Technical Specifications
This experimental equipment prohibits: Testing and storage of samples of flammable, explosive, and volatile substances Testing and storage of corrosive substance samples Testing or storage of biological samples Testing and Storage of Strong Electromagnetic Emission Source Samples | |||
Equipment model |
LX-408L |
||
Studio Size |
Length x Width x Height 800 x 600 x 850mm |
||
External dimensions |
Length x Width x Height 1350 x 1670 x 1280mm |
||
total power |
About 12KW |
||
Main performance parameters |
temperature range |
-40℃~+150℃ |
|
Temperature deviation |
≤±2.0℃ |
||
temperature fluctuation |
≤±1℃ |
||
Humidity range |
30~98% R.H |
||
Humidity deviation |
≤±5%R.H |
||
Humidity fluctuation |
≤±2.5%R.H |
||
Heating and cooling rate |
The average cooling rate is 0.7 ℃ to 1.5 ℃/min; The average heating rate is 3 ℃/min |
||
electrical control system |
|
||
display |
Equipment |
||
7 "color LCD display |
Operation mode |
||
Fixed value or program |
Setting method |
||
Chinese/English menu, touch screen input |
Program capacity |
||
Can store 50 sets and 50 segments of test procedures |
program length |
||
Each group of programs can have a maximum of 50 segments |
Number of cycles |
||
Each program segment can loop 999 times |
Setting Range |
||
Temperature: -40 ℃ to+150 ℃ | |||
Humidity: 0-99.9% RH | |||
Time: 0-99 hours and 59 minutes |
distinguish |
||
Rate | |||
Temperature: 0.1 ℃ | |||
Humidity: 0.1% RH |
Time: 0.1min |
||
input |
PT100 platinum resistor |
||
Auxiliary functions |
Alarm display function (indicating the cause of the fault), power outage protection function, upper and lower limit temperature and humidity alarm function, timing function (automatic start and stop), self diagnostic function. |
||
Measurement data collection |
PT100 platinum resistor |
Cabinet material |
|
Inner box material |
SUS304 stainless steel plate |
||
Outer box material |
Cold rolled steel plate spray coating |
||
thermal insulating material |
Rigid polyurethane foam+glass fiber |
||
door |
Single door, equipped with observation window, lighting, left hinge, left handle |
Heating and humidification system heater |
|
Nickel chromium alloy electric wire heater |
Heater control method: PID control method, using contactless equal period pulse width modulation SSR (solid-state relay) humidifier Shallow groove electric heating humidification method Stainless steel armored electric heater |
||
Humidifier control method: PID control method, using contactless equal period pulse width modulation SSR (solid-state relay) |
Water level control device, heater anti dry burning device |
refrigeration system |
|
refrigeration compressor |
French Taikang fully enclosed piston compressor |
||
Cooling method |
Mechanical compression unit refrigeration |
||
Cooling method |
forced air cooling |
||
evaporator |
Light tube heat exchanger |
||
evaporative condenser |
Brazed plate heat exchanger |
||
condenser |
air-cooled condenser Control mode of refrigeration machine The PLC (Programmable Logic Controller) of the control system automatically selects and adjusts the operating conditions of the refrigeration machine according to the test conditions Condensation pressure regulating valve Evaporator Pressure Regulator |
||
Compressor return air cooling circuit |
Energy regulation circuit |
||
refrigerant |
R404a/R23 |
Safety device |
|
Leakage protector |
Compressor overpressure protector |
||
Over temperature protector inside the box |
Compressor overheat protector |
||
Fan overheat protector |
Compressor overcurrent protector |
||
phase sequence relay |
Waterway system protector |
Water supply system |
|
|
Water supply method |
Pump lifting |
|
|
Water storage tank |
1 built-in water tank |
|
water quality |
Clean and purified water (clean tap water is also acceptable) |
||
Standard configuration | |||
1 hollow electric tempered glass observation window (W350mm * H280mm) | |||
1 fluorescent lamp (AC220V) | |||
1 cable hole (Φ 50mm) with rubber plug | |||
Power cord 5m |
4 sample shelves and 2 shelves |
Conditions of Use power supply AC380V three-phase four wire+protective earth wire; ○ Voltage allowable fluctuation range: AC380V ± 10% |
|
○ Allowable frequency fluctuation range: 50Hz ± 0.5 Hz |
The grounding resistance of the protective grounding wire is less than 4 Ω; TN-S power supply or TT power supply |
||
Usage environment |
Temperature: 5 ℃~35 ℃, relative humidity: ≤ 85% RH site ○ The ground is flat, well ventilated, and free of flammable, explosive, corrosive gases and dust |
||
There is no strong electromagnetic radiation source nearby |
○ Leave appropriate maintenance space around the equipment |
||
Meet standards | |||
GB10586-200 Technical Conditions for Damp Heat Test Chamber | |||
GB10589-2006 Technical Conditions for Low Temperature Test Chamber | |||
GB10592-2006 Technical Conditions for High and Low Temperature Test Chambers | |||
GB11158-2006 Technical Conditions for High Temperature Test Chamber | |||
GB/T2423.1-2001 Test A: Low temperature test method | |||
GB/T2423.2-2001 Test B: High temperature test method | |||
GB/T2423.3-1993 Test Ca: Constant Damp Heat Test | |||
GB/T2423.4-1993 Test Db: Alternating Damp Heat Test Method | |||
| |||
GB/T5170.2-1996 Basic parameter calibration methods for environmental testing equipment for electrical and electronic products Temperature testing equipment |
GB/T5170.5-1996 Basic parameter calibration methods for environmental testing equipment for electrical and electronic products - Damp heat testing equipment after-sale service |
||
The equipment warranty period is three guarantees and one year, with lifelong maintenance. We offer a one-year warranty from the date of delivery (except for damages caused by natural disasters, abnormal electricity, improper use, or improper maintenance, our company provides free repairs). For services that exceed the warranty period, corresponding cost fees will be charged. | |||
Any problems that arise during the use of the equipment shall be responded to within 24 hours, and maintenance engineers and technicians shall be promptly dispatched to handle and solve the problems. |
The supplier provides free training for the buyer's operators and maintenance personnel, usually 2-3 people.
When the supplier's equipment malfunctions after the warranty period, the supplier provides paid services. (Fees may be charged at discretion) |
Configuration List |
Serial Number |
name |
1 |
Origin brand |
|
notes |
2 |
controller |
5166 7 inches |
Electromagnetic switch |
3 |
Shilin, Taiwan |
SP-11 |
Intermediate relay |
4 |
Omron Japan |
MY2J |
timer |
5 |
Japan CKC |
AH3-2 |
over temperature protection |
6 |
South Korea RAINBOW |
TS-320S |
SSR |
7 |
Yangming, Taiwan |
ASR03-40DA |
temperature sensor |
8 |
Baoguang |
PT100 |
light |
9 |
domestic |
220V18W |
Non fuse switch |
10 |
Shilin, Taiwan |
220V40A |
circulating fan |
11 |
YEZHENG |
220V60W |
wind wheel |
12 |
Taiwan Shangyu |
6F |
compressor |
13 |
Taikang, France |
4P |
|
14 |
condenser |
Taiwan Science and Technology Power |
Condenser fan |
15 |
Germany Dunly |
220V350M |
Dry filter |
16 |
Danfoss Denmark |
DCL0325 |
One-way valve |
17 |
Danfoss Denmark |
NRV16S |
solenoid valve |
19 |
Japanese Egret Worker |
NEV202DXF |
evaporator |
20 |
Hongjian |
customized |
refrigerant |
21 |
Dupont |
R404A/R23 |
heater |
22 |
Taiwan Weide |
220V2000W |
water pump |
24 |
Karen |
500ml/min |
power supply |