main features
Measure 4 parameters: vapor pressure, temperature, atmospheric pressure, relative humidity
Compatible with SDI-12, can use SDI-12 communication protocol, convenient for accessing data acquisition from manufacturers other than METER
Compact in size, with sensors measuring only 5 cm x 2 cm
Good waterproof performance, unique Teflon surface makes ATMOS 14 able to withstand rainstorm
Fast response speed, PTFE surface is resistant to liquid water and dust, and gaseous water can penetrate well, so it can quickly respond to humidity changes
Synchronized monitoring of temperature and humidity for higher accuracy
Equipped with radiation shielding cover
TECHNICAL INDEX
relative humidity
Range: 0-100% RH
Resolution: 0.1% RH
Accuracy: varies due to different relative humidity, as shown in Figure 1
Response time (τ, 63%):<40s (when the airflow velocity is 1m/s)
Hysteresis: Typical<1% RH
Long term drift: typical<0.5% RH/year
Figure 1 Relative humidity accuracy
temperature
Scope:- 40~ 80℃
Resolution: 0.1 ℃
Accuracy: varies depending on temperature, as shown in Figure 2
Response time (τ, 63%):<400s (when the airflow velocity is 1m/s)
Long term drift: typical<0.04 ℃/year
Figure 2. Temperature accuracy
vapor pressure
Range: 0~47 kPa
Resolution: 0.001 kPa
Accuracy: Different due to temperature and humidity, as shown in Figure 3
Figure 3. Accuracy of water vapor pressure
atmospheric pressure
Range: 50~110 kPa
Resolution: 0.01 kPa
Accuracy: ± 0.4 kPa
output: DDI serial port or SDI-12
size: 2.0 cm(dia)×5.4 cm(h)
operation temperature:- 40~80℃
interface type: 3.5mm headphone jack or tin plated bare wire jack
Cable lengthStandard 5 meters, customizable in other sizes up to 75 meters in length
Support data collection (to be purchased separately):METER data acquisition (ZL6, EM50/60 series); Other data acquisition devices that can collect 3.6-15.0 VDC signals and communicate through serial ports or SDI-12 communication
Supply voltage (VCC to GND): 3.6~15.0VDC
Digital input voltageLogic high 2.8~5.5V, typical 3.0V; Logic low -0.3~0.8V, typical 0.0V
Digital output voltageLogic High Typical 3.6V
Power line conversion rate: Minimum 1.0V/ms
current consumptionDuring the measurement period, 4.50~5.00 mA, typical 4.75 mA; Typical 0.03mA during static period
TimeTypical 550ms; Maximum 600ms
Start Time: DDI serial port maximum 100ms; SDI-12 100~1100ms, Typical 1100ms
Capacity per meter of cableTypical 147 pF
Resistance per meter of cableTypical 34 m Ω
Following standardsProduced under the ISO 9001:2015 standard; 2004/108/EC and 2011/65/EU; EN61326-1:2013; EN50581:2012
Origin and Manufacturer:AmericaMETERCompany(Former Decagon)