Jinan Jingce Electronic Technology Co., Ltd
Home>Products>Atmospheric detection chromatograph
Atmospheric detection chromatograph
The detection range of atmospheric detection chromatograph: indoor gases and total volatile organic compounds (VOCs) in the atmosphere, acetone in exh
Product details

Atmospheric detection chromatographMain performance characteristics:

1. The instrument adopts a fully microcomputer automatic control system, an adjustable LCD display screen, and rich and intuitive display content; Chinese keyboard style operation, setting parameters and operation are very convenient.

2. Adopting a brand new dual CPU design and original Japanese high integration electronic circuit technology, it has excellent reliability and anti-interference ability. The temperature control accuracy reaches 0.1 ℃, and the display accuracy is as high as 0.01 ℃. It can also display the power consumption percentage of each heating part at the same time.

3. The design of a three gas path and a large volume column chamber allows for the simultaneous installation of three chromatography columns, FID and TCD detectors, and facilitates the expansion of ECD, FPD, and NPD detectors, as well as the installation of external control equipment such as gas injectors and converters.

4. This machine can be easily expanded from a single FID amplifier to a dual FID amplifier, truly achieving multi-purpose use of one machine.

5. Equipped with two packed column vaporization chambers and an independent capillary dedicated system, the design of the vaporization chamber draws on the miniaturization design of similar foreign products, with fast heating speed, no injection dead corners, excellent capillary diversion/non diversion injector and tail blowing system, making it easy to install small-diameter capillary columns, large-diameter capillary columns, and various packed columns.

6. It has a five order linear program heating function, which is suitable for multi-component sample analysis.

7. The column chamber is equipped with a frequency conversion intelligent rear door automatic cooling system, which has a very fast cooling speed (from 280 ℃ to 50 ℃ in just 5 minutes), achieving true near room temperature operation.

8. Equipped with six independent temperature control systems Suitable for the installation of multiple temperature control requirements such as conversion furnaces; The column chamber, gasification chamber, hydrogen flame detection chamber, thermal conductivity detection chamber, auxiliary 1, and auxiliary 2 all adopt modular design, making installation and replacement more convenient.

9. It has power failure protection function and can automatically memorize set parameters. It has anti power mutation interference function and stopwatch timing function.

10. Equipped with over temperature protection function, the power automatically shuts off and an alarm is triggered when the temperature exceeds the set upper limit. If the temperature of any circuit exceeds the set temperature, heating will automatically stop; And there are prompts for the cause of the malfunction and alarm prompts.

11. It has a "0 ℃" protection function. When the temperature display of any channel is always "0 ℃", the heating will automatically stop to prevent the heating element from being damaged due to the constant heating of the chamber And there are prompts for the cause of the fault and alarm prompts

12. Among similar products, the gas cut-off protection and prompt function can maximize the protection of TCD tungsten wire and chromatographic column from damage.

13. It has a wide range of fault self diagnosis functions, 36 types of self diagnosis information, and can display the cause and handling methods of faults at any time.

14. Stable gas path system: (1) The carrier gas flow rate adopts a dual stability design of first stabilizing the pressure and then stabilizing the flow; (2) Accurate pneumatic control valve components with high reproducibility and stability.

15. Multiple injection methods are available: (1) filling column head injection; (2) Filling column gasification injection; (3) Capillary diversion/non diversion injection: back pressure valve control technology, linear diversion; (4) Six way valve gas injection;

(5) Automatic headspace injection;

Temperature control indicators: 1) Column box: Room temperature+5 ℃ -400 ℃ Accuracy: 0.1 ℃ Display accuracy up to 0.01 ℃ 2) Sample injector: Room temperature+5 ℃ -400 ℃ Accuracy: 0.1 ℃ Display accuracy up to 0.01 ℃ 3) Detector: Room temperature+5 ℃ -400 ℃ Accuracy: 0.1 ℃ Display accuracy up to 0.01 ℃ 4) Column temperature box volume: 248 * 248 * 260MM5) Program heating order: Fifth order linear rise 6) Program heating rate: 0.1 ℃ -40 ℃ (0.1 ℃ increment) 7) Constant temperature holding time for each order: 0~9999.9min (0.1min increment)


Atmospheric detection chromatographTechnical specifications (configured as FID, optional for other detectors, up to 3 detectors can be installed on one instrument):

detector

Sensitivity/Sensitivity

noise

drift

linear

characteristic

FID

Mf ≤ 1 × 10-11 g/s (benzene)

≤ 0.02mV

≤ 0.15mV/h

≥107

Circular collection electrode, corundum ceramic nozzle

Note: Dimensions: 660 * 460 * 465mm Weight: 60kg Input power supply: AC 220V ± 10% 50HZ 2KW

Attachment 1:

The most commonly used detectors in gas chromatographs nowadays are:

1. TCD detector (thermal conductivity cell detector) is the most widely used detector, suitable for inorganic gases and organic compounds, and is particularly suitable for constant analysis or analysis of components above one hundred thousandth

2. FID detector (hydrogen flame ionization detector) has high sensitivity and is suitable for trace analysis of organic compounds

3. ECD detector (electron capture detector)

It is a low-energy radioactive detector widely used in pesticide, air, and water pollution detection, suitable for analyzing organic complexes of halides, polysulfides, polycyclic aromatic hydrocarbons, and metal ions.

4. FPD detector (flame photometric detector)

Suitable for analyzing pesticides containing sulfur and phosphorus, as well as monitoring trace amounts of organic pollutants containing sulfur and phosphorus in environmental analysis

5. NPD detector (nitrogen phosphorus detector)

It is a quality detector suitable for analyzing nitrogen and phosphorus compounds with high sensitivity and selectivity; It can detect 5 × 10 (-13 times) g/s azobenzene nitrogen-containing compounds and 2.5 × 10 (-13 times) g/s phosphorus containing compounds, such as marathon pesticides. It has a high response to nitrogen and phosphorus compounds. And for other compounds, some response values are 10000 to 100000 times lower.