BINDER KMF115 Constant Temperature and Humidity Chamber
1、 Product Introduction:
1. Temperature range: -10 ℃ to 100 ℃
2. Larger humidity range: 10% relative humidity to 98% relative humidity
3 APT.lineTMPreheating chamber technology
4. Using capacitive humidity sensors and steam humidification for humidity regulation
5. The interior is entirely made of stainless steel
6. BINDER communication software APT-COM ™ 3 Basic versions
7. Intuitive touch screen controller using interval programming and real-time programming
8. Internal data logger and open format measurement values can be read through USB
9. Equipment self-test is used for comprehensive status analysis
10. The sealed inner door is made of safety glass (ESG)
11. Prevent corrosion of glass through a special TIMELESS coating.
12. 1 socket made of stainless steel
13. Access hole with silicone plug 30mm, left
14. Independently adjustable temperature safety device Class 3.1 (DIN 12880), equipped with optical and acoustic temperature alarms
15. Computer interface: Ethernet
16. Door heating
2、 Technical parameters:
PRODUCT MODEL | KMF115-230V1 | KMF115-240V1 |
Choose a model | standard | standard |
Item code | 9020-0341 | 9020-0342 |
measure | ||
Internal volume [L] | 102 | 102 |
Net weight of equipment (unloaded) [kg] | 127 | 127 |
Maximum total load [kg] | 100 | 100 |
Maximum load per partition [kg] | 30 | 30 |
The dimensions of the casing do not include additional components and connectors | ||
Net weight width [mm] | 880 | 880 |
Net weight at height [mm] | 1050 | 1050 |
Depth net weight [mm] | 650 | 650 |
Wall to side distance [mm] | 100 | 100 |
Wall distance from rear [mm] | 100 | 100 |
Internal dimensions | ||
Width [mm] | 600 | 600 |
Height [mm] | 483 | 483 |
Depth [mm] | 351 | 351 |
The number of doors | ||
Inner door | 1 | 1 |
Outer door | 1 | 1 |
Data related to the environment | ||
Energy consumption at 85 ℃ and 85% relative humidity [Wh/h] | 570 | 570 |
Sound pressure level [dB (A)] | 52 | 52 |
firmware | ||
Number of partitions (hours/maximum) | 1/5 | 1/5 |
Temperature performance data | ||
Temperature range [℃] | -10…100 | -10…100 |
The average heating rate is based on IEC 60068-3-5 [K/min] | 1.3 | 1.3 |
The average cooling rate is based on IEC 60068-3-5 [K/min] | 0.5 | 0.5 |
The maximum thermal compensation is at 25 ℃ [W] | 150 | 150 |
Climate performance data | ||
Temperature range [℃] | 10…90 | 10…90 |
The temperature deviation depends on the set value [± K] | 0.2…1 | 0.2…1 |
The temperature fluctuation depends on the set value [± K] | 0.1…0.3 | 0.1…0.3 |
Humidity range [% relative humidity] | 10…98 | 10…98 |
The humidity fluctuation depends on the set value | ≤ 2.5 ± relative humidity | ≤ 2.5 ± relative humidity |
Condensation temperature range of condensate water [℃] | 5…90 | 5…90 |
Electric data | ||
Rated voltage [V] | 200…230 | 200…240 |
Power frequency [Hz] | 50/60 | 50/60 |
Rated power [kW] | 2 | 2 |
Equipment safety device [A] | 16 | 16 |
Phase (rated voltage) | 1~ | 1~ |
1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. The technical data assumes a fan speed of 100%. We reserve the right to make technical changes.