BINDER MKF 240 High and Low Temperature Alternating Climate Chamber
1、 Product Introduction:
1. Temperature range: -40 ℃ to 180 ℃
2. Humidity range: 10% relative humidity to 98% relative humidity
3. Integrated water storage tank 20 L
4. Four zero voltage switch contacts that can be activated through the MCS controller
5 APT.lineTMPreheating chamber technology
6. Programmable condensation protection for samples
7. Heated observation window with LED internal lighting
8. Using capacitive humidity sensors and steam humidification for humidity regulation
9. BINDER communication software APT-COM 3 Basic version
10. Fault diagnosis system using optical and sound alarms
11. Controller using interval programming
12. Access hole with silicone plug 50mm, left
13. Independent adjustable temperature safety device level 2 (DIN 12880) with optical alarm
14. 4 casters, 2 with brakes
15. Computer interface: Ethernet
16. 230 V socket on the right side operating component
17. Display through LCD color monitor
18. Adjustable slope function function function
19. Integrated continuous automatic recorder
20. Real time clock
21. Door heating
22. Alarm message when fresh water tank is short of water
23. One socket made of stainless steel
24. Hose protection joint, including water supply and drainage pipes with a height not exceeding 1 meter
25. BINDER testing certificate
26. Inner chamber made of stainless steel
27. PT100 temperature sensor
28. Heat insulation through polyurethane foam
29. Fluorine free refrigerant R-404A
30. Cooling through a compression cooler
31. Finned evaporator
2、 Technical parameters:
PRODUCT MODEL | MKF240-400V1 | MKF240-400V-C1 |
Choose a model | standard | Equipped with transformer and frequency converter |
Item code | 9020-0285 | 9020-0295 |
measure | ||
Internal volume [L] | 228 | 228 |
Net weight of equipment (unloaded) [kg] | 360 | 360 |
Maximum total load [kg] | 70 | 70 |
Maximum load per partition [kg] | 30 | 30 |
The dimensions of the casing do not include additional components and connectors | ||
Net weight width [mm] | 1115 | 1115 |
Net weight at height [mm] | 1715 | 1715 |
Depth net weight [mm] | 925 | 925 |
Wall to side distance [mm] | 300 | 300 |
Wall distance from rear [mm] | 235 | 235 |
Observation window width [mm] | 508 | 508 |
Observation window height [mm] | 300 | 300 |
Internal dimensions | ||
Width [mm] | 735 | 735 |
Height [mm] | 700 | 700 |
Depth [mm] | 443 | 443 |
The number of doors | ||
Outer door | 1 | 1 |
Data related to the environment | ||
Sound pressure level [dB (A)] | 65 | 67 |
firmware | ||
Number of partitions (hours/maximum) | 1/6 | 1/6 |
Temperature performance data | ||
Temperature range [℃] | -40…180 | -40…180 |
The temperature deviation depends on the set value [± K] | 0.1…1.5 | 0.1…1.5 |
The temperature fluctuation depends on the set value [± K] | 0.1…0.5 | 0.1…0.5 |
The average heating rate is based on IEC 60068-3-5 [K/min] | 5 | 5 |
The average cooling rate is based on IEC 60068-3-5 [K/min] | 5 | 5 |
The maximum thermal compensation is at 25 ℃ and 90% relative humidity [W] | 400 | 400 |
The maximum thermal compensation is at 25 ℃ [W] | 2800 | 2800 |
Climate performance data | ||
Temperature range [℃] | 10…95 | 10…95 |
The temperature fluctuation depends on the set value [± K] | 0.1…1.3 | 0.1…1.3 |
Humidity range [% relative humidity] | 10…98 | 10…98 |
The humidity fluctuation depends on the set value | ≤ 2.5 ± relative humidity | ≤ 2.5 ± relative humidity |
Condensation temperature range of condensate water [℃] | 5…94 | 5…94 |
Electric data | ||
Rated voltage [V] | 400 | 480 |
Power frequency [Hz] | 50 | 60 |
Rated power [kW] | 6.8 | 6.8 |
Equipment safety device [A] | 16 | 16 |
Phase (rated voltage) | 3~ | 3~ |
1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. The technical data assumes a fan speed of 100%. We reserve the right to make technical changes.