Shanghai Fenglin Electronic Technology Co., Ltd
Home>Products>Bacterial Turbidity Meter WGZ-XT-MA Turbidity Meter
Bacterial Turbidity Meter WGZ-XT-MA Turbidity Meter
Bacterial Turbidity Meter WGZ-XT-MA Turbidity Meter
Product details

  bacteriaturbidimetersummary:WGZ-XTBacterial turbidity meterMainly used to determine the bacterial concentration in the suspension of the strain to be identified. This instrument is calibrated using (BaSO4) McFarland turbidity standard solution, using MCF (McFarland) McFarland turbidity units.

The bacterial turbidity meter directly displays the turbidity value in units of wheat. Suitable for bacterial liquid concentration determination in medical and health institutions, biological products, quarantine institutions, and scientific research institutions.

 细菌浊度仪WGZ-XT,麦氏浊度
Bacterial turbidity meter, MacLehose turbidity meter

Principle of Bacterial Turbidity Meter:

The basic principle of a bacterial turbidity meter is that after the measuring device is placed in the sample, the light beam emitted by the light source forms scattered light when it encounters bacterial particles. The resulting 90 ° scattered light of the turbidity signal is received by a photosensitive element, and the light signal is amplified by a circuit and processed by a single-chip microcomputer to display the measured turbidity value.

Bacterial turbidity meter button instructions:

Correction key: Enter the correction state and perform zero position correction on blank values

Confirm button: Confirm the previous operation

Storage key: Save measurement values

Query key: Query saved measurement values

4、 Characteristics of bacterial turbidity meter:

1. Microcomputer, touch keyboard, LCD backlit display screen, making reading more comfortable and not affected by natural light

2. Quickly set the average measurement mode to obtain accurate data in the shortest possible time..

3. Simple operation, appropriate measurement range, and high cost-effectiveness make it more suitable for use in various industries.

4. Unique positioning structure and high-precision optical path system effectively ensure the correctness and repeatability of measurement values.

5. Adopting a low drift, high-precision, and stable circuit system, as well as an efficient and long-life light source, can effectively ensure the long-term stable operation of the instrument

6. Can store 20 historical records without losing them during power outages

5、 Technical parameters and indicators of bacterial turbidity meter

PRODUCT MODEL

WGZ-XT

Minimum indication

0.001 MCF

measuring range

0-6MCF (Mercantile Turbidity Units)

1MCF=3×108CFU/ml

Indication error

±6%(±5%F. S)

repeatability

≤0.5%

zero drift

±0.5%F.S

Product Features

Bacterial turbidity analysis, direct reading (MCF) McPherson turbidity units.

Working conditions of bacterial turbidity meter

1. The instrument should be placed in a dry room and used at a temperature of 5 ℃ -35 ℃.

2. When using, it should be placed on a flat workbench and avoid vibration.

3. Indoor lighting should not be too strong, and direct sunlight should be avoided.

4. Try to stay away from high-intensity magnetic fields, electric fields, and electrical equipment that generate high-frequency waves.

5. Avoid high temperatures near the instrument.

6. Power supply for the instrument: 220V ± 22V, 50 ± 1Hz, and must be equipped with a good grounding wire.

Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!