KD-2040TypeCODAmmonia nitrogen and total phosphorus analyzer
1Instrument features
1.Separating the digestion instrument from the measuring instrument does not affect the measurement. temperaturePIDAutomatic temperature control and timing.
2.Time saving operation.CODDissolve only requires15min.
3.Cold light source, narrowband interference optical system, good optical stability.
4.Equipped with calibration tube function to ensure measurement accuracy.
5.Can be saved separatelyCODStandard curves for ammonia nitrogen and total phosphorus10Article and999The measured value is not lost when the power is turned off.
6.LCDLarge screen LCD display, easy and intuitive operation.
7.It has printing function (printer is an optional accessory).
8.The host chassis is molded by compression moldingABSThe material has good corrosion resistance.
IITechnical indicators
1.Measurement range:
COD:5~2000mg/L
(Divided into three ranges: 5~200 mg/L、200~1000 mg/L、1000~2000 mg/L).
Ammonia nitrogen:0.02~25mg/L
Total phosphorus:0.00~10mg/L
2.Indication error:
COD:The indication error is ≤±5%
Ammonia nitrogen: ≤±3%(F.S)
Total phosphorus:≤±5%(F.S)
3.Repeatability:≤3%
4.Anti chlorine interference: ≤2000mg/L(CODMeasurement)
5.Temperature control system: room temperature~180℃Can be set,CODThe digestion temperature is165℃, total phosphorus digestion temperature 120 ℃.
6.Temperature control accuracy:±0.5℃
7.Temperature control time:1~999minAdjustable
8.Dissolving time:CODby15minThe total phosphorus is30 min
9.Optical system:610nm(COD)The420nm(CODAmmonia nitrogen)630 nm(Total phosphorus)
10.Optical stability: The absorbance value of the instrument is within20minInternal drift less than0.002A
11.Batch processing capacity:16A water sample
12.External dimensions:CODAmmonia nitrogen total phosphorus analyzer 266mm×200mm×130mm
Dissolving instrument(DIS-16Type)216mm×320mm×146mm
13.Weight: host1kgDissolver6kg
14.power waste:<500W
1.Normal usage conditions:
⑴ Environmental temperature: 5~40℃
⑵ Relative humidity: ≤ 85%
⑶ Power supply: AC(220±22)V; (50±0.5)Hz
⑷ No significant vibration or electromagnetic interference, avoid direct sunlight.
IIIConfiguration List
host1Taiwan, digestion instrument1Taiwan, digestion colorimetric tube30Support, test tube rack1One,
CODAmmonia nitrogen and total phosphorus1Reagent kit, digestion protective cover1One,
an instruction manual1Copy, Product Qualification Certificate1Copy and warranty card1Share.