CU-600 Total Organic Carbon Analyzerprinciple
Using ultraviolet (UV) to oxidize slow-moving water samples, organic matter in the samples is oxidized to CO2, CO2The measurement adopts conductivity detection technology to obtain the TOC value of the water sample by measuring the change in conductivity of the water sample before and after oxidation.
CU-600 Total Organic Carbon AnalyzerTechnical Parameter
TOC detection range: 0.1 µg/L~1500.0 µg/L
TOCDetection accuracy: 0.1 µ g/L (conductivity<1.0 µ S/cm)
Detectable items: TOC、 Conductivity and temperature
Conductivity detection range: 0.055-6.000 µ S/cm
TOC accuracy error: ≤ ±5%
Repetitive error: ≤ 3%
Analysis time: The shortest analysis interval of 20 seconds (can be set by oneself)
Response time: within 15 minutes
Sample temperature: 0-100 ℃
Zero drift: ± 2%/D
Range drift: ± 2%/D
Injection method: peristaltic pump
Environmental temperature: 10-40 ℃ Temperature variation within ± 5 ℃/d
Relative humidity: ≤ 85%
Power Requirements: 220V AC ≤± 10% 50Hz
Power: 60W
Basic dimensions: 30cm × 20cm × 18cm
Weight: 10kg
CU-600characteristic
- High detection accuracy, low detection limit
- Touch screen design, user-friendly interface, simple and convenient operation
- Conductivity temperature compensation technology to improve detection accuracy
- No need for carrier gas and reagents, simple maintenance, low cost
- A separate system suitability test interface for more user-friendly operation
- Real time visible detection curve, more intuitive
- Large data storage capacity, easy to query, and can be queried on a daily basis
- Built in needle printer reduces space occupation
- Countdown of UV lamp lifespan for easy replacement and maintenance of UV lamps
- Can be connected externally with 4-20mA
- Equipped with alarm output function beyond the upper limit
- Optional online sampling device for online monitoring of water system
- Optional automatic calibration package, one click completion of automatic calibration and system applicability testing
- Optional automatic sampler for unmanned operation, saving manpower and time
CU-600Application scope
Offline laboratory testing of injection water and purified water in the pharmaceutical industry;
TOC online monitoring of water systems in the pharmaceutical industry;
Ultra pure water preparation system and chip process in the semiconductor industry;
Process of preparing deionized water for power plants