Working Principle
1The instrument adopts the principle of chemiluminescence detection. The sample undergoes high-temperature oxidation reaction, and the nitrogen compounds in it are quantitatively converted into NO2. The sample gas is dehydrated by a membrane dryer and enters the reaction chamber. The metastable NO reacts with O3 from the ozone generator in the reaction chamber and is converted into excited NO2. When the excited state of NO2 transitions to the ground state, photons are emitted, and the light signal is detected and received by a photomultiplier tube at a specific wavelength. By amplifying microcurrents and processing computer data, it can be converted into an electrical signal proportional to the intensity of light. The intensity of chemiluminescence is proportional to NO, and the amount of NO generated is proportional to the total nitrogen content in the sample. Therefore, the total nitrogen content in the sample can be determined by measuring the intensity of chemiluminescence.
2Execution standards: GB/T 17674、SH/T 0657、ASTM D5453-9
Product Overview
1、 The GWM-N200 chemiluminescence nitrogen analyzer is currently the most advanced nitrogen element synchronous analyzer in China, widely used to detect nitrogen content in liquid, solid, and gas samples.
2The instrument uses chemiluminescence to determine the total nitrogen content, and the key components of the system are imported, ensuring reliable performance of the entire machine.
3The instrument is suitable for measuring the total nitrogen content of paraffin oil, diesel, gasoline, lubricating oil, fuel oil, liquefied gas and natural gas, as well as other oil products, chemical raw materials and finished products.
Technical Parameters:
project |
total nitrogen determination |
||
Various types of samples |
liquid |
solid |
gas |
Measurement method |
Chemiluminescence method (N) |
||
Sample injection volume |
1~50uL |
1~100mg |
1~10mL |
measuring range |
0.2~10000mg/L |
||
measurement accuracy |
Concentration value (ppm) |
Injection volume (uL) |
RSD(%) |
0.2 |
20 |
25 |
|
5 |
10 |
10 |
|
50 |
10 |
8 |
|
100 |
10 |
5 |
|
More than 5000 |
5 |
3 |
|
Temperature control range and accuracy |
Room temperature~1100 ℃, ± 1 ℃ |
||
Air source requirements |
Argon: purity above 99.995%, oxygen: purity above 99.99% |
||
Power requirements |
AC220V±22V, 50Hz±0.5Hz,1500W |
||
Instrument dimensions |
660×510×420 |
||
Instrument weight |
35Kg(Net weight 30kg) |