Jiangsu Hualu Instrument Co., Ltd
Home>Products>Current and voltage calibrator
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
    18015194111
  • Address
    No. 201 Jinhe Road, Jinhu County, Huai'an City, Jiangsu Province
Contact Now
Current and voltage calibrator
Current and voltage calibrator
Product details

Detailed Introduction

Current and voltage calibratorIt is a handheld comprehensive digital verification instrument that integrates digital display DC voltage, current signal source, and DC current functions, with high precision, high resolution, high reliability, and anti drop performance. The current and voltage calibrator adopts a large LCD display with clear readings. At the same time, the instrument also has backlight, reading retention function, and automatic shutdown function. The instrument can be powered by DC or AC power, making it more convenient to use.
List of Output and Measurement Functions


function

measure

output

notes

DC Voltage

0~30 V

0~10 V

DC millivolt

0~100 mV

0~100 mV

dc current

0~24 mA

0~24 mA

frequency

1.000Hz~99.999kHz

0.00Hz~20.000kHz

other

24V power supply, step output, slope output, step output

DC voltage measurement


range

*Large measurement range

Resolution

Accuracy (reading accuracy)%+Counting)

SK-1815 (0.05 level)

SK-2015 (0.02 level)

30V

0V~31V

0.001V

0.05+2

0.02+2

100mV

-15mV~80mV

0.001mV

0.05+20

0.02+20

80mV~125mV

0.01mV

0.05+2

0.02+2

-The temperature coefficient for temperatures ranging from 10 ℃ to 18 ℃ and+28 ℃ to 55 ℃ is ± 0.005% FS/℃.
Input resistance:> 1M Ω.

DC voltage output


range

*Large output range

Resolution

Accuracy (reading accuracy)%+Counting)

SK-1815 (0.05 level)

SK-2015 (0.02 level)

100mV

-15mV~99.999mV

0.001mV

0.05+20

0.02+20

100mV~125mV

0.01mV

0.05+2

0.02+2

10V

0~11V

0.001V

0.05+2

0.02+2

-The temperature coefficient for temperatures ranging from 10 ℃ to 18 ℃ and+28 ℃ to 55 ℃ is ± 0.005% FS/℃.
*Large load: 1mA or 1k Ω, depending on the lighter load.

DC current measurement


range

*Large measurement range

Resolution

Accuracy (reading accuracy)%+Counting)

SK-1815 (0.05 level)

SK-2015 (0.02 level)

20mA

0~24mA

0.001mA

0.05+2

0.02+2

-The temperature coefficient for temperatures ranging from 10 ℃ to 18 ℃ and+28 ℃ to 55 ℃ is ± 0.005% FS/℃.
Input resistance: < 100 Ω.

DC current output


range

*Large output range

Resolution

Accuracy (% of reading)+Counting)

SK-1815 (0.05 level)

SK-2015 (0.02 level)

20mA

0~24mA

0.001mA

0.05+2

0.02+2

20mA
(Transmitter simulation)

0~24mA

0.001mA

0.05+2

0.02+2

-The temperature coefficient for temperatures ranging from 10 ℃ to 18 ℃ and+28 ℃ to 55 ℃ is ± 0.005% FS/℃.
*High load voltage: 20V, equivalent to the voltage of 20mA current on a 1000 Ω load resistor.

frequency


range

*Large measurement range

Resolution

Accuracy (reading accuracy)%+Counting)

SK-1815 (0.05 level)

SK-2015 (0.02 level)

100Hz

1~99.999Hz

0.001Hz

0.02+1

0.01+1

1000Hz

100~999.99Hz

0.01Hz

10kHz

1k~9.9999kHz

0.1Hz

100kHz

10k~99.999kHz

1Hz

Sensitivity: 10Hz to 10kHz, Vp-p≥1V; The remaining Vp-p is ≥ 2V.
Waveform: Square wave, additional 5 counts are required for errors in other waveforms, which can directly measure the mains frequency.

Frequency output


range

*Large output range

Resolution

Accuracy (reading accuracy)%+Counting)

SK-1815 (0.05 level)

SK-2015 (0.02 level)

200Hz

0~200Hz

0.01Hz

0.02+1

0.01+1

2000Hz

0~2000Hz

0.1Hz

20kHz

0~20kHz

1Hz

Output amplitude: ≥ 4.5Vp-p;
Waveform: Square wave.


Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!