This instrument can achieve chemical analysis and digestion of COD, total phosphorus, total nitrogen, and total chromium in various water samples such as surface water, industrial wastewater, and domestic sewage that require heating processes. 16 reaction tubes with a diameter of 16mm can be placed simultaneously, and the temperature can be automatically controlled according to the set digestion temperature and time.
Dissolving instrument | |||
Dissolving temperature: | Room temperature~200 ℃, automatic constant temperature control | Temperature control accuracy: | ±0.5℃ |
Dissolving time: | 1~9999min | Quantity of single digestion samples: | 16 of them |
Timing accuracy: | 0.1 seconds per hour | Automatic shutdown: | Settable |
monitor: | 3.5-inch color touch screen with a resolution of 320x240, displaying in both Chinese and English | Power Supply: | AC220V ± 10%, 50Hz, power 500W (maximum) |
Weight: | 5.5kg | Shell: | anti-corrosion |
External dimensions: | 355mm × 260mm × 135mm (excluding protective cover) |
High temperature control accuracy. With multi-point error free calibration, zero error measurement of the set point is achieved, completely eliminating the errors caused by the nonlinearity of temperature sensors in common digesters on the market. Effective insulation measures have been adopted to reduce energy consumption. It has a timed reminder and can be set to automatically shut down after digestion. Chinese English display, color touch screen. The instrument has a secondary protection function. When the temperature of the digestion instrument exceeds 200 degrees, the instrument will cut off the power by itself to ensure the safety of the digestion process. The standard digestion tube is made of imported glass and has the characteristic of not bursting during rapid cooling and heating. |
name | quantity | name | quantity |
host | 1 unit | shield | 1 piece |
Dissolving tube | 16 pieces | power line | 1 piece |