Dalong MicroPette Plus Fully Disinfected Manual (Adjustable&Fixed) Pipette
Dalong MicroPettePlus fully disinfected manual fixed pipette can disinfect the entire pipette under high temperature and pressure conditions of 121 ℃
Product details

Dalong MicroPette PlusFully disinfected manually&Fixed typePipette

The fully disinfected pipette can be used1211atmUnder conditions of high temperature and high pressure disinfection of the entire unit20Minutes, no special preparation is needed to disinfect the entire tube. After disinfection, it must be placed over12Allow it to cool and dry thoroughly within hours.It is recommended to check the performance of the pipette after each round of high-temperature and high-pressure disinfection. It is also recommended that every10Re lubricate the piston and sealing ring of the pipette after sub high temperature and high pressure disinfection.

Characteristics of single channel pipette
-
Lightweight and designed to conform to ergonomics
-
Digital window, making the set range clear at a glance
-
Wide range of measurement,0.1μlto10ml
-
Using attachment tools allows for convenient and efficient calibration and maintenance
-
Accurate liquid separation, each pipette is operated according toEN/ISO8655Standard calibration
-
The entire unit can withstand high temperature and high pressure disinfection
-
Adopting new materials and high-temperature and high-pressure resistant structures

The multi-channel pipette also has the following innovative features
-
The nozzle ejector can simultaneously eject multiple suction nozzles, which is efficient and labor-saving
-
Various ranges of8Dao and12The pipette is suitable for standard use96perforated plate
-
Liquid head can360Rotating degrees, convenient for pipetting
-
Each nozzle connection has an independent piston device, making maintenance and repair convenient
-
Special nozzle connector design, easy to observe the sealing condition of the suction nozzle


Ordering Information

MicroPette plusSingle channel manually adjustable pipette

Item number

range

increment

Measure volume(μl

Allow * * system error
(Inaccuracy)

Allow * * random error
(Inaccurate)





%

μl

%

μl

0.1-2.5μl

0.05μl

2.5

2.50%

0.0625

2%

0.05

1.25

3%

0.0375

3%

0.0375

0.25

12%

0.03

6%

0.015

0.5-10μl

0.1μl

10

1%

0.1

0.80%

0.08

5

1.50%

0.075

1.50%

0.075

1

2.50%

0.025

1.50%

0.015

2-20μl

0.5μl

20

0.90%

0.18

0.40%

0.08

10

1.20%

0.12

1%

0.1

2

3%

0.06

2%

0.04

5-50μl

0.5μl

50

0.60%

0.3

0.30%

0.15

25

0.90%

0.225

0.60%

0.15

5

2%

0.1

2%

0.1

10-100μl

1μl

100

0.80%

0.8

0.15%

0.15

50

1%

0.5

0.40%

0.2

10

3%

0.3

1.50%

0.15

20-200μl

1μl

200

0.60%

1.2

0.15%

0.3

100

0.80%

0.8

0.30%

0.3

20

3%

0.6

1%

0.2

50-200μl

1μl

200

0.60%

1.2

0.15%

0.3

100

0.80%

0.8

0.30%

0.3

50

1%

0.5

0.40%

0.2

100-1000μl

5μl

1000

0.60%

6

0.20%

2

500

0.70%

3.5

0.25%

1.25

100

2%

2

0.70%

0.7

200-1000μl

5μl

1000

0.60%

6

0.20%

2

500

0.70%

3.5

0.25%

1.25

200

0.90%

1.8

0.30%

0.6

1000-5000μl

50μl

5000

0.50%

25

0.15%

7.5

2500

0.60%

15

0.30%

7.5

1000

0.70%

7

0.30%

3

2-10ml

0.1ml

10ml

0.60%

60

0.20%

20

5ml

1.20%

60

0.30%

15

2ml

3%

60

0.60%

12

8Dao manual adjustable pipette

Item number

range

increment

Measure volume(μl

Allow * * system error
(Inaccuracy)

Allow * * random error
(Inaccurate)






%

μl

%

μl


0.5-10μl

0.1μl

10

1.50%

0.15

1.50%

0.15


5

2.50%

0.125

2.50%

0.125


1

4%

0.04

4%

0.04


5-50μl

0.5μl

50

1%

0.5

0.50%

0.25


25

1.50%

0.375

1%

0.25


5

3%

0.15

2%

0.1


50-300μl

5μl

300

0.70%

2.1

0.25%

0.75


150

1%

1.5

0.50%

0.75


50

1.50%

0.75

0.80%

0.4


12Dao manual adjustable pipette

Item number

range

increment

Measure volume(μl

Allow * * system error
(Inaccuracy)

Allow * * random error
(Inaccurate)






%

μl

%

μl


0.5-10μl

0.1μl

10

1.50%

0.15

1.50%

0.15


5

2.50%

0.125

2.50%

0.125


1

4%

0.04

4%

0.04


5-50μl

0.5μl

50

1%

0.5

0.50%

0.25


25

1.50%

0.375

1%

0.25


5

3%

0.15

2%

0.1


50-300μl

5μl

300

0.70%

2.1

0.25%

0.75


150

1%

1.5

0.50%

0.75


50

1.50%

0.75

0.80%

0.4


Manually fix the pipette

Item number

range

increment

Measure volume(μl

Allow * * system error
(Inaccuracy)

Allow * * random error
(Inaccurate)






%

μl

%

μl


5μl

-

5μl

1.3%

0.065

1.2%

0.06


10μl

-

10μl

0.8%

0.08

0.8%

0.08


20μl

-

20μl

0.6%

0.12

0.5%

0.1


25μl

-

25μl

0.5%

0.125

0.3%

0.075


50μl

-

50μl

0.5%

0.25

0.3%

0.15


100μl

-

100μl

0.5%

0.5

0.3%

0.3


200μl

-

200μl

0.4%

0.8

0.2%

0.4


250μl

--

250μl

0.4%

1.0

0.2%

0.5


500μl

-

500μl

0.3%

1.5

0.2%

1.0


1000μl

-

1000μl

0.3%

3.0

0.2%

2.0


2000μl

-

2000μl

0.3%

6.0

0.15%

3.0


5000μl

-

5000μl

0.3%

15

0.15%

7.5