Electronic load displacement measuring instrumentpurpose
Electronic load displacement measuring instrumentAn instrument suitable for measuring the load characteristics and displacement changes of cement electric poles.
2. After configuring different force sensors, the mechanical performance testing of all rod-shaped material products can be expanded, and it can also be tested in all situations of other loads and displacements.
A display instrument has2A display window that can be configured with 2 force sensors and1A displacement sensor.Low tension display0.01KN,Displacement measurement display1mm. Can display clearly under strong light.
Indicator functional conditions
1. Purpose: Measurement1Test within 00KN
Measurement: Winding type 18Displacement measurement within 00mm
(Display unit 1mm)
Force display unit
1At 00KN The reading is 0.1KN
1At 0KN The reading is 0.01KN
3. Displacement display unit in millimeters
4. Measurement accuracy ≤1% . Reading ± 1 word
5. Force display adjustmentzeroZero adjustment
6. Zero adjustment of displacement display
7. Environmental temperature for use -5℃ ~+45℃
8. Use environmental humidity 80% (relative humidity)
9. Power supply voltage range~220 ± 15% (50Hz)
10. There is no strong magnetic field around during use.
11. Allowable load error: 1%
usage method:
1. First, connect the selected load sensor to the "Force Signal Input" socket on the left side of the rear panel of the tester using a five core cable signal line10KN or1After zeroing, it displays 000KN
2. TakeTwist the displacement sensor and displacement signal line. The other end is connected to the socket on the back panel (displacement input) of the tester.
3. Connect the power socket to the rear panel of the tester(220V Power Supply)Plug in socket, withConnect the 220V power supply.
4. Press the front panel (power switch) to turn on the light, and the LCD will display numbers.
5.At this time, turn the zero knob of the two windows and adjust them to000
The 6.00mm displacement sensor should display 000mm
7.Load window display readingThe 10KN sensor is 00.000KN
There should be a 15 minute warm-up process from startup to operation, and work should begin after the numbers have stabilized.
. Storage and maintenance
1. The digital load displacement tester is a precision electronic instrument. Before use, it is necessary to have a thorough understanding of the instrument and be familiar with the testing methods.
When the instrument is not in use, the sensor should be removed and packed together with other accessories in the packaging box.
3. The packaging box should be placed in a dry and oil-free place.
4. This instrument should be handled gently Handle with care Prevent strong vibrations.
5. The fuse is located on the back panel of the tester, in the power socket (or fuse holder), and can be opened with a small Phillips knife to a capacity of 1A.
6. This instrument must not be opened by itself to avoid affecting the testing accuracy.
When users discover a malfunction in the display instrument, they can request to send it to a designated repair station or manufacturing unit for repair.