Beijing Tiannuojie Technology Co., Ltd
Home>Products>G2307 gas concentration analyzer
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
    13501352085
  • Address
    No. 5 Changchun Bridge Road, Haidian District, Beijing, China
Contact Now
G2307 gas concentration analyzer
G2307 gas concentration analyzer
Product details

Product Overview

The Picarro G2307 gas concentration analyzer can accurately measure formaldehyde (H2CO), methane (CH4), and water vapor (H2O) in real-time. This analyzer is equipped with coating components in key gas pathways, which reduces the tendency of H2CO molecules to adhere to the surface of the gas pathway, thereby improving the measurement response time. This analyzer is an ideal solution for atmospheric research and monitoring of formaldehyde in automobile and aircraft exhaust, photochemical smog, and incinerator emissions. It can be used to measure sensitive trace amounts of formaldehyde and environmental formaldehyde to ensure indoor air quality.

Product Features

Fast, continuous, and real-time measurement of H2CO

Long term stability to meet low-frequency calibration requirements

Small footprint, suitable for on-site or laboratory deployment, no need for consumables

Equipped with water correction function, automatically reporting dry gas mole fraction

Main technical parameters

Picarro G2307performance index

H2CO

CH4

H2O

Detection lower limit(3σ, 300second)

0.3 ppb

6 ppb

-

Zero drift(24hour)

(peak-peak value, fiftyMinute average)

1.5 ppb

-

-

accuracy(1σ, 2second)accuracy(1σ, 10second)accuracy(1σ, 300second)

1.2 ppb + 0.1%The reading

0.6 ppb + 0.05%The reading

0.1 ppb + 0.02%The reading

20 ppb + 0.2%The reading10 ppb + 0.1%The reading

2 ppb + 0.05%The reading

10 ppm + 0.1%The reading

-

-

Measurement interval

<2second

-

-

Accuracy

±10%

±2%

±5%

response time(0-20 ppb)

Descent time90–10% : < oneMinute rise time10–90% : < oneminute

-

-

measuring range

0-30 ppm

0-20 ppm

0-3%

Picarro G2307System indicators

measuring technique

Cavity ring down spectroscopy(CRDS)

Sample temperature

-10 to 45centigrade

Sample flow rate

~ 400Standard cubic centimeters per minute, in760Under atmospheric pressure, no filtration is required

Sample pressure

300 reach 1000 Tuo(40reach 133 Kilopascals)

Sample humidity

<99%Relative humidity, in40Under non condensing conditions at degrees Celsius, drying is not necessary

ambient temperature

10 to 35Celsius (during instrument operation);-10 to 50Celsius (instrument storage conditions)

ambient humidity

<99%Relative humidity, non condensing conditions

parts

Vacuum pump (external, included), keyboard included, mouse included, LCD display (optional)

data output

RS-232Interface, network interface,USBInterface, analog signal (optional)0–10 V

Pipe joint

¼inchPFA Swagelok ®

size

Analyzer:17Inches widex 7Inch heightx 17.5Inch deep(43.2 x 17.9 x 44.6centimeter), not included0.5Inch foot pad external pump:7.5Inches widex 4Inch heightx 11Inch deep(19 x 10.2 x 28centimeter)

Installation form

Workbench (standard) or19Inch rack mounted chassis (optional)

weight

59.3 pound(26.9Kilogram), including external pumps

Electricity demand

100 – 240Alternating current,47 – 63Hertz (automatic detection),<260Total power of watt startup;

110 Watt (analyzer)+ 80Watt (vacuum pump), when operating stably.

Application precautions

  When the concentration of organic compounds in the sample far exceeds the normal environmental level, it may interfere with the measurement, including but not limited to ethane, acetylene, ammonia, and other nitrogen-containing and sulfur-containing compounds. Users should verify whether the experimental samples are suitable. If unsure, please contact us to discuss the specific situation of the experiment.

Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!