GGX-820 atomic absorption spectrophotometer
Application areas: food hygiene, geology, metallurgy, agricultural testing, chemical environmental protection, water quality monitoring, pharmaceutica
Product details

Product Features
1. Integrated design of power supply host.
2. Fast heating rate: ≥ 2000℃/s

3. It has multiple safety monitoring measures:

(1) Temperature monitoring of graphite furnace body

(2) Gas pressure monitoring

(3) Monitoring of graphite tube fracture or lack of clamping

(4) Transformer temperature monitoring
3. Graphite furnace automatic sampler ensures high-precision positioning (0.1%) of the graphite furnace injection needle for injection accuracy.

4. Equipped with a 52 position automatic sampler.


optical system
C-T monochromator with 1800 grating lines/mm, focal length of 270mm, shining wavelength of 250nm, short optical path, suitable for element measurement in the ultraviolet region.
Spectral bandwidth automatic switching, available in four levels: 0.2, 0.4, 1.0, and 2.0nm.


Atomization system
Vertical heating of graphite tube.
Temperature control range: room temperature~3000 ℃; Temperature control error (2000 ℃) ≤± 0.5%.
Heating rate ≥ 2000 ℃/s, maximum 10 step heating program, with three heating modes: step, ramp, and hold.


Graphite furnace automatic sampler
Sample size: 52 bits (including 3 5mL positions)
Injection volume: 2 μ L-100 μ L
Injection accuracy (20 μ L): ≤± 0.5%
Can automatically configure the working curve and support the addition of matrix modifiers at will


Safety function
Internal and external gas pressure monitoring
Over temperature protection for power transformers
Over temperature protection for the furnace
Graphite furnace fracture alarm



Wavelength Range

190-900nm

wavelength accuracy

±0.1nm

baseline drift

±0.003A/30min

Spectral bandwidth deviation

±0.02nm

Characteristic Concentration

Cu<0.02μg/mL/1%

Background correction capability

≥60times

detection limit

Graphite furnace (Cd), Q L: ≤ 0.5pg

Measurement accuracy(RSD

Graphite furnace (Cd): ≤ 2.0%

Wavelength Repeatability

≤0.1nm


Requirements for Instrument Working Environment
Power supply: 220VAC ± 10%, 50Hz ± 1Hz
Temperature: 15 ℃ -35 ℃
Relative humidity: 10% -80%


Patent technology:

Zeeman background correction optical path and system ZL 2015100677596

A multi position rotating lighthouse lighting system ZL 2018207884083


Technical parameters of the whole machine

Optical platform type: floating
Background correction method: deuterium lamp clip background
Instrument size: (L × H × D): 870mm×510mm×525mm
Power: Host 200VA, Graphite Furnace: Instantaneous Maximum Power 4000VA