Detailed description of gas bath constant temperature oscillator
Characteristics of gas bath constant temperature oscillator:
The gas bath constant temperature oscillator, also known as the constant temperature shaker, uses a pre control system and LED display for temperature control in this series of products. High temperature control accuracy, convenient temperature adjustment, accurate and intuitive indication, superior and reliable performance. The gas bath constant temperature oscillator, incubator, and studio are equipped with lighting devices for easy observation. The oscillating incubator, also known as the full temperature oscillator, uses a high-quality fully enclosed compressor with a large cooling capacity. The incubator is equipped with fans and devices to force air convection, resulting in a more uniform temperature distribution.
Application of gas bath constant temperature oscillator:
The air bath constant temperature oscillator is suitable for environmental protection, medical teaching, health and epidemic prevention, drug testing, animal botany, marine science and food engineering, and other scientific research and production departments. It is a specialized constant temperature equipment with an oscillator for BOD determination in water analysis, cultivation and preservation of bacteria, viruses, molds, and microorganisms, plant cultivation, and breeding experiments.
Main technical parameters of gas bath constant temperature oscillator:
name | model | Temperature Control Range | Temperature fluctuation | Power | rotational speed | Studio Size |
Air bath Constant temperature oscillator | CHA-SA THZ-82A | RT—50℃ | ±1℃ | 600W | 30-300rpm/min | 580×490×240 |
Main technical parameters and product features of gas bath constant temperature oscillator:
The air bath constant temperature shaker (also known as the air constant temperature shaker) is a biochemical instrument that combines a temperature controlled constant temperature incubator and an shaker. It is mainly suitable for the oscillation cultivation of various liquid and solid compounds such as biology, biochemistry, cells, and bacterial strains in various universities, medical, petrochemical, health and epidemic prevention, environmental monitoring and other scientific research departments. Its main features are: A: Accurate numerical display of temperature control. B: Provide oxygen replenishment holes and a constant temperature working chamber with sufficient oxygen replenishment. C: Equipped with mechanical timing. D: The universal spring bottle rack is particularly suitable for the cultivation and preparation of biological samples for various comparative experiments. E: Stepless speed regulation, smooth operation, easy and safe operation. F: The inner cavity is made of stainless steel, which has good corrosion resistance.
Main technical performance of gas bath constant temperature oscillator:
1、 Using power supply: 220V 50Hz
2、 Heating power: 400w
3、 Timing range: 0-120 minutes (or normally open)
4、 Oscillation frequency: starting -300 revolutions per minute, adjustable
5、 Oscillation amplitude: 20mm
6、 Constant temperature range: room temperature -50 ℃
7、 Oscillatory methods: reciprocating, swirling, and dual function (to be selected during procurement)
8: Temperature control accuracy: 1℃
9: Special reminder: Add 'A' after the model to indicate temperature Speed dual digital display
Notice before using the gas bath constant temperature oscillator:
1. Please read this manual carefully before use.
2. The working surface of the machine should not be excessively smooth (such as tiles, etc.).
3. The electrical rated parameters of the socket provided by the user should not be less than the electrical rated parameters of the machine, and there should be good grounding measures.
4. The whole machine is strictly prohibited from being used in an environment with direct sunlight.
5. The speed control should start slowly from low speed to high speed.
6. The machine may experience some displacement during high-speed oscillation, so it should be monitored by someone during use.
Instructions for using the gas bath constant temperature oscillator:
1. Load the test bottle and maintain balance. If it is a dual function model, set the oscillation mode. (For detailed settings, please refer to the following diagram)
2. Connect the power and set the timer according to the surface scale of the machine. If you need to work for a long time, adjust the timer to the "normally open" position.
3. Turn on the power switch and set the constant temperature:
(1) Place the control switch in the 'Setting' section, and the temperature displayed on the screen will be the set temperature. Adjust the knob to set it to the desired temperature for your work. (The working temperature you set should be higher than the ambient temperature. At this point, the machine will start heating up and the yellow indicator light will be on. Otherwise, the machine will not work.)
(2) Place the small switch of the control part on the 'measurement' end, and the temperature displayed on the screen will be the actual temperature of the air inside the test chamber. As the temperature inside the chamber changes, the displayed number will also change accordingly.
(3) When heated to the desired temperature, the heating will automatically stop and the green indicator light will turn on; When the heat inside the test chamber dissipates below the temperature you set, a new round of heating will begin.
4. Activate the oscillation device:
(1) Turn on the oscillation switch on the control panel, and the indicator light will turn on.
(2) Adjust the oscillation speed knob to the desired oscillation frequency.
5. Cut off the power after work, and set the speed control knob and temperature control knob to the lowest point.
6. Clean the machine and keep it clean.
Precautions for gas bath constant temperature oscillator:
1. Instruments should be placed on a sturdy workbench, with a clean and tidy environment and good ventilation.
2. The power socket provided by the user should have good grounding measures.
3. It is strictly prohibited to move the machine during normal operation.
4. It is strictly prohibited for objects to collide with the machine.
5. Children are strictly prohibited from approaching the machine to prevent accidents.
6. Before replacing the fuse, ensure that the power supply has been cut off.
7. After use, please clean the machine without leaving any water droplets or dirt residue.
Gas bath constant temperature oscillator can be customized according to customer requirements
After sales service of gas bath constant temperature oscillator: three guarantees for one year, lifelong maintenance!