CHARACTERISTICS
photoLab S6belt6An interference filter and circular colorimetric dish slot, suitable for routine detection of sewage and tap water.photoLab S12belt12An interference filter. Square colorimetric dishes and large packaged reagents can be used to save analysis costs. In drinking water testing, trace level accuracy can be achieved.
1AQA/IQCquality assurance
2Automatic recognition of colorimetric tubes
3Bar code reader
4AutoSelectAutomatically select the corresponding testing method without the need for user selection
5MemoChipMaking instrument upgrades very easy without the need for replacement EPROM
6The numeric keypad allows users to easily change setting values and labels(photoLabS12)
7AutoselectorAutomatic selector is used to automatically select user-defined testing methods(photoLab S12)
8The usable specifications are10、20、50mmThe square colorimetric dish(photoLab S12)
IITechnical Parameter
model |
photoLab S6 |
photoLab S12 |
category |
automatic photometer |
automatic photometer |
band |
6 A(330,445,525,550,605,690) |
12 A(340,410,445,500,525,550,565,605,620,665,690,820) |
accuracy |
± 2 nm |
± 2 nm |
half-width |
10 nm |
10 nm |
AUTO ZERO |
have |
have |
AutoSelect |
have |
have |
Automatic confirmation of colorimetric tube |
have |
have |
Colorimetric tube |
cylindrical |
cylindrical; square 10,20,50mm |
rechargeable battery |
have |
have |
data storage |
500 Group, with date and time |
1000 Group, with date and time |
interface |
RS232 |
RS232 |
Battery operation status |
One day, discharge protection, automatic charging when using a transformer as a power source |
One day, discharge protection, automatic charging when using a transformer as a power source |
date/time |
Built in real-time clock |
Built in real-time clock |
size |
140 × 270 × 260 mm |
140 × 270 × 260 mm |
weight |
2.3kg(With battery2.8kg) |
2.3kg(With battery2.8kg) |
authentication |
CE |
CE |