Jinan Jingce Electronic Technology Co., Ltd
Home>Products>Graphite furnace atomic absorption spectrometer
Graphite furnace atomic absorption spectrometer
The graphite furnace atomic absorption spectrometer complies with the analysis methods for various metal element contents in the 2010 edition of the P
Product details

SDA-100FGHGraphite furnace atomic absorption spectrometer

1、 Performance indicators

1. Optical system
Automatic adjustment of wavelength range 190nm-900nm
Grating engraved line density 1800 lines
Monochromator aberration correction Czerny Turner type monochromator
Spectral bandwidth of 0.1, 0.2, 0.4, 1.0, and 2.0nm, with multiple automatic switching modes
Wavelength accuracy ± 0.20nm
Wavelength repeatability of 0.1nm
Baseline drift static 0.005A/30min dynamic 0.008A/30min
Detector wide spectral range photomultiplier tube
The resolution can distinguish the dual lines of 279.5nm and 279.8nm in the manganese lamp, and the energy of the valley between the two spectral lines should be less than 20%


2. Light source system
Lamp holder 3 lamp holder
Lamp power supply method: 400Hz square wave pulse
Lamp current regulation range 0-10mA average current
3. Atomization system
Flame analysis method
Detection limit (Cu) 0.006 μ g/ml
Burner 100mm all titanium metal burner
Precision RSD ≤ 1%
Spray metal sheathed high-efficiency glass spray
Corrosion resistant all plastic atomization chamber
The safety measures have multiple automatic safety protection functions, including acetylene leakage alarm and automatic shutdown system
Flameless (graphite furnace) analysis method
Detection limit (Cd) ≤ 4pg

4. Flame analysis method and measurement method

Air acetylene flame method, hydride generator atomic absorption method
The concentration calculation method automatically fits the working curve, automatically corrects sensitivity, and calculates concentration and content.
Repeat measurements 1 to 99 times, automatically calculate the mean, content, standard deviation, and relative standard deviation.
Print test data or final analysis report as a result. The results can be exported as an Excel worksheet.
Software environment: Windows XP/Vista operating system, professional Chinese software

5. Flameless (graphite furnace) analysis method

Measurement method standard addition method, interpolation method
Background correction: Deuterium lamp background correction is greater than 30 times, and self absorption buckle background correction is greater than 60 times
Repeat the measurement 1-3 times, calculate the average, and provide the standard deviation and relative standard deviation
Print test data or final analysis report as a result. The results can be exported as an Excel worksheet.
Software environment: Windows XP/Vista operating system, professional Chinese software

IIParameters of graphite furnace atomic absorption spectrometer

Power supply AC 220V/50HZ
Power 150VA
Net weight of the instrument 55kg
Instrument gross weight 80kg
Instrument volume 1110mm (horizontal) x 400mm (depth) x 500mm (height)
Packaging volume 1350mm (horizontal) x 600mm (depth) x 850mm (height)
Working environment temperature: 10-35 ℃
Working environment humidity ≤ 80%


IIIStandard configuration
1. Host
2. Low noise oil-free air compressor
3. Three hollow cathode lamps
4. Operating software

4Design Features
1. Total reflection optical system with chromatic aberration reduction across the entire wavelength range. After adjusting the focal point, any wavelength can be changed, and the focal point will no longer change. The wavelength will be automatically scanned and peak searched.


2. It has a deuterium lamp buckle background and a self-priming buckle background.

3. The uniquely designed all titanium burner ensures smooth combustion and fast thermal equilibrium speed.

4. Automatically adjust negative high voltage, lamp current, and element lamp position to maximize energy, and provide automatic gas leakage and flameout protection. 3 hollow cathode lamp stands. Automatic control system, capable of automatically scanning wavelengths. Preheat one element lamp and preheat the other two element lamps.

5. Graphite furnace heating: Users can freely set the heating curve according to their needs.

Introduction to hydride generator:

Hydrogen Generator

Hydrogen generator: a supporting equipment for atomic absorption spectrophotometer and spectrometer, used to detect the content of some low melting point elements (such as lead, arsenic, mercury, etc.)

Essential analytical equipment for environmental monitoring, food hygiene, quality inspection, commercial inspection, and drinking water testing.
High degree of automation, only one key press is needed to complete the entire process of injection, reaction, reading, and cleaning.
By adopting advanced flow injection technology, the reaction of the sample is more uniform, which improves the stability of the equipment operation,

Relative standard deviation RSD<2%.
Measurement rate: 100 times/hour
Single sampling volume: less than 2ml

Typical element sensitivity

element

wavelength

Typical sensitivity

The best sensitivity

As

193.7

0.15

0.08

Se

196.0

0.30

0.12

Sb

217.6

0.30

0.15

Bi

223.0

0.35

0.18

Sn

286.3

0.35

0.12

Pb

283.3

0.15

0.08

Te

214.3

0.35

0.12

Hg

253.7

0.50

0.35

Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!