Shanghai Huishi Instrument Equipment Co., Ltd
Home>Products>HSLC-100 Fully Automatic Distillation Range Tester
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
    18501761259
  • Address
    East Building, Blue Treasure Building, No. 29 Jiatai Road, Shanghai Pilot Free Trade Zone, China
Contact Now
HSLC-100 Fully Automatic Distillation Range Tester
HSLC-100 Fully Automatic Distillation Range Tester
Product details



According to the GB/T6536-2010 standard, the instrument allows users to choose automatic pressure calibration or not to calibrate the pressure. The system is equipped with an atmospheric pressure measuring device, which has high measurement accuracy. At the same time, the instrument also comes with fully automatic monitoring of temperature, air pressure, auxiliary equipment, fire extinguishing equipment, and liquid level tracking equipment. Once the system is not working properly, it will automatically give a prompt and take corresponding measures immediately to prevent accidents from occurring.
The instrument has the following characteristics:
1. Compact structure, beautiful appearance, and easy operation.
2. Adopting the principle of fuzzy temperature control, the temperature control accuracy is high and the response speed is fast.
3 10.4-inch ultra large true color touch screen makes use simple.
4. High level tracking accuracy.
5 fully automatic distillation range process, fully automatic process monitoring.
Product parameters:
1. Temperature measurement range: Room temperature to 400 ℃
Temperature control accuracy: ± 0.1 ℃
3 Steam temperature sensor: PT100 platinum resistor
Heating from 4 to initial boiling point time: 5-15 minutes
5 Distillation rate (flow rate from initial boiling point to 95% recovery): 4-5ml/min
The time from the residual amount of 5ml in the 6-flask to the endpoint is no more than 5 minutes
The initial boiling point temperature is detected by an optical fiber sensor
8 Recycling volume measurement range: 0-100ml
9 Recycling volume measurement accuracy: ± 0.1 ml
10 volume measurement: liquid level tracking system, mainly composed of stepper motor and optical fiber liquid level sensor, the motor steps 0.01ml each time
11 Cooling System: Compressor Refrigeration
12 Cold bath temperature: 0-60 ℃
13 Recycling Room Temperature: 10-40 ℃
The 14 devices have the function of automatically detecting local atmospheric pressure and automatically correcting the temperature to standard atmospheric pressure
15 comes with a fire extinguishing system, which can be manually extinguished with nitrogen gas in case of a fire to prevent safety accidents caused by the rupture of the heating distillation flask (nitrogen cylinders need to be provided by the customer)
16 comes with a safety temperature setting function, allowing room temperature to be freely set between 400 ℃. If the heating temperature exceeds the set safety temperature, the instrument can automatically stop the experiment and immediately blow air to dissipate heat.
This instrument can automatically detect and display the temperature and flow rate time at every 5% volume point.
The 18 instrument can automatically detect the temperature and occurrence time of the initial boiling point and final boiling point.
19. Multiple automatic stop modes can be selected, such as automatic stop at the final boiling point, stop when the specified temperature is reached, and stop when the specified volume is reached.
Online inquiry