HXD-1000TMS/LCD automatic turret HV, HK dual head digital microhardness tester
HXD-1000TM/LCD Automatic Rotary Tower Digital Microhardness Tester HXD-1000TMS/LCD Automatic Rotary Tower HV, HK Dual Head Digital Microhardness Teste
Product details

HXD-1000TM/LCDAutomatic turret digital microhardness tester
HXD-1000TMS/LCDAutomatic turret HV, HKDual head digital microhardness tester
Digiital Micro Hardness Tester
 

HXD-1000TMS/LCD自动转塔HV、HK双压头数显显微硬度计


Innovative design of operation control keys and LCD display screen. Designed as a menu style structure, it can display indentation diagonals D1 and D2, hardness scale HV or HK, hardness value, test force, holding time, measurement frequency, Vickers Rockwell conversion value, date and time, measurement statistics, and loading process on different pages.

In the system settings menu, the test force unit (gf or N), LCD backlight, date, and time can be set separately. By touching the keys, the brightness of the light source can be infinitely adjusted, the tower can be rotated, and the loading can be controlled. The instrument also has a power-off memory function and a timing device for measuring positions. Through the printer, measurement data such as maximum value, minimum value, uniformity, mean value, and mean square error can be output.

Novel design of control buttons and LCD display. Auto motorized tureet.

With the touch buttons and the LCD screen, it is easy and visual to set hardness scale (HV or HK), date; to select the force unit (gf or N), force duration; to adjust the background of LCD and the brightness of illuminating lamp; to operate the turret; to display and output diagonal lengths D1-D2, measured HV or HK hardness values, statistics (Max, Min, Ave, deviation), converted HRC HRA values, and etc.

technical specifications

model
Type
HXD-1000TM/LCD
HXD-1000TMS/LCD
objective/Pressure head switching
Lens/Indenter switch
Automatic turret switching
With Motorized Turret
Lens/HV/HK Indenter switch with Motorized Turre
Test force N
Test force
9.807, 4.903, 2.942, 1.961, 0.9807, 0.4903, 0.2452, 0.0981 N
1000, 500, 300, 200, 100, 50, 25, 10 gf
Loading Method
Loading mode
Programmable automatic test force loadingMaintain and uninstall
Program-controlled application, duration and removal of load
Dwell time
Duration time
5 – 60 sec selectable
Step size of 5 seconds
Hardness measurement range
Measuring range
5 HV -- 3000 HV
Measuring microscope
Microscope
objectiveObjective: 40x;
10X (for observation)
Eyepiece: 15x
Total magnification: 600x
150X (for observation)
Minimum testing unit
Min. detection
0.025μm
X-Y test bench
X-Y stage
Dimensions: 100X100mm
Maximum travel range: 25X25 mm
Minimum Reading Resolution: 1/100mm
Optical path switching
Optical path
Measurement/CCD photography switching
Measurement / Photo switch
Measurement reading method
Measuring mode
Digital display for eyepiece measurement
With eyepiece digital display

data output
Data output
  Digital display of test force, holding time, hardness value, average value, non-uniformity, standard deviation, Rockwell HRC HRA conversion value, etc. The built-in printer provides test reports.
Display on LCD: Load, Duration time, Diagonal lengths, Hardness values, Min, Max, Average, Divergence,, Standard deviations, converted HRC HRA values, and etc. A test result is provided with the built-in printer.
 
Maximum height of test piece
Max..test height
85mm
Execution standards
Standard
Execution standard: National Standard of the People's Republic of China
GB/T 4340.2-1999 Metallic Vickers Hardness Tester Test
Part 2: Inspection of Hardness Tester
Shanghai Enterprise Standard Q/NWFQ 01-2002 Microhardness Tester

Standard accessories, attachments
Standard attachments

Objective 40x, 10x, Vickers indenter (on board)
Measure eyepiece 15X standard hardness block (700HV)1400HV0.2)
Thin film clamp, small cylinder clamp, flat clamp, level gauge
Halogen bulb power cord fuse auxiliary tool
User manual printer (built-in)
Digital Microscopic Camera Eyepiece and Software SDE-100
Optional parts
Options
Knoop indenter
Digital camera and adapter
Power supply
220 VAC, 50/60 Hz
Dimensions Outsourcing size 54 x 50 x 64 cm (Packing size)
Gross weight 47 kg
 
 
Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!