PRODUCT INFORMATION
measuring range:0.00 to 10.00 mg/L(ppm)CI2
Adopting innovative optical system design, it has excellent accuracy, repeatability, and measurement time required,HannauniqueCAL Check ™The function allows for performance validation and calibration of instruments using NIST traceable standards, with built-in reaction timers to ensure consistency between sample measurements and user results, and good laboratory practices(GLP)Display the date and time of the last user calibration, battery status indicator light, and automatic shutdown settings.
reminder:NEWHI97734Residual chlorine total chlorine analyzer,Adopting a menu operation mode, advanced optical system, narrowband interference filter can provide more accurate and repeatable readings, a tutorial mode with simple step-by-step instructions, and automatic recording of recent data50Secondary measurement
TECHNICAL INDEX
technical specifications |
HI96734Residual chlorine [free chlorine] - total chlorine |
measuring range |
0.00 to 10.00 mg/L(ppm)CI2 |
resolution |
0.01 mg/L (≤ 3.50ppm), 0.10 mg/L (>3.50ppm) |
accuracy@25°C (77°F) |
Reading±3% ± 0.03 mg/L(ppm) |
optical system |
narrow band interference filter,silicon photocellTungsten lamp @ 525 nm |
Method standards |
USEPAmethod330.5, DPDmethod |
Power Mode |
1 x 9V battery |
Automatic shutdown |
In measurement mode, measurement mode10In minutes, or in calibration mode, it will automatically shut down after one hour of no use |
application environment |
0 to 50 °C(32 to 122°F ); RH-max95%(No condensation) |
Dimension |
Host size:192 x 104 x 69 mm (7.6 x 4.1 x 2.7 ")Host weight:320 g (11.3 oz.) |
Exclusive attachments |
HI731333Glass cuvette(HI731331+HI731335)Specification: Outer diameter24.6Mm (inner diameter)22mm),4Set/Group |
HI93734-01 Customized residual chlorine (free chlorine) - total chlorine DPD reagent prediction times100order |
|
|
HI93734-03 Customized residual chlorine (free chlorine) - total chlorine DPD reagent prediction times300order |
HI96734-11 total chlorineNISTMeasurement curve calibration group, standard value@25ºC:0.0 and 4.00 ± 0.20 mg / L(ppm)CI2 |
Standard configuration
HI96734