High purity chromatography gas source oil-free air generator
High purity chromatography gas source oil-free air generatorWorking principle:
The gas chromatography air generator is powered by a fully enclosed imported compressor, which purifies natural air through four stages to remove moisture, oil, and impurities from the air, and outputs stable and clean air through a stabilizing device.
Product advantages:
The air generator has the characteristics of stable output pressure, low noise, and strong purification ability, making it an ideal air source to replace high-pressure air cylinders. This product can not only meet the needs of various models of domestic and imported chromatographs and various analytical laboratories, but also serve as an air source for high-purity nitrogen generators.
Instrument features:
- Air generators replace high-pressure steel cylinders to make laboratories more instrument based.
- The air generator has a small volume, light weight, reasonable structure, simple operation, and convenient use.
- The air generator has low vibration, low noise, stable output pressure, and air pressure fluctuations not exceeding ± 0.003Mpa.
- The air generator is safe and reliable, equipped with secondary protection devices for overvoltage and overcurrent, low pressure starting, and effectively improves the service life of the compressor.
- The air generator has a long service life, can be used continuously or intermittently, has stable performance, and does not degrade flow rate and purity.
- The instrument is equipped with an HL-A device, which automatically adsorbs and removes hydrocarbons, esters, oils, etc. from the air, greatly improving the quality of the air without maintenance.
- The generator adopts stainless steel gas storage tank and automatic water discharge; (To prevent the discharge of yellow rust from contaminating the laboratory).
- The gas circuit of the generator is entirely made of stainless steel pipes (electrolytic polishing, ultrasonic cleaning).
- The generator has dual filters, two-stage voltage regulation, high output pressure accuracy, and complete protection measures.
- The air generator adopts advanced silent and oil-free compressors, which have the characteristics of oil-free, low noise, stable output pressure, easy and safe operation, novel appearance design, small volume, and light weight. It is an ideal air source for gas chromatography, instrument analysis, and laboratory replacement of steel cylinders.
Main technical parameters of gas generator:
- Air purity: oil-free grade 3
- Output flow rate: 2L/min 5L/min
- Output pressure: A: 0-0.4Mpa B: 0-0.5Mpa
- Working conditions: Power supply voltage: 220V ± 10%
Environmental temperature: 10-43 ℃
Relative humidity: 85%
There is no significant dust pollution or corrosive gases in the environment
- Work noise:<40 decibels
- Instrument power: 150W
- Safety valve opening pressure: 0.85 ± 0.01Mpa
® Shanghai Xuansheng Science