Japanese Tuji GL100 Multi functional Portable Data Recorder Communication Function: The GL100-N standard data collector can communicate with a computer through a USB interface to read data from the Japanese Tuji GL100 multifunctional portable data logger. It can also control the GL100-N to start or stop recording, change testing conditions, etc. GL100-N provides dedicated analysis software for free. The wireless version of the Japanese Tuji GL100 multifunctional portable data recorder can achieve wireless communication with a computer or GL240/GL840 through WIFI signals. The computer or GL240/GL840 recorder can remotely read data from the Japanese Tuji GL100 wl. The Japanese graphic technology GL100-wl can also be remotely operated through a computer.

content Real time viewing of temperature and humidity data is possible. The free software for GL100-N/GL100-WL mini temperature and humidity recorder can communicate with a PC computer through a randomly provided USB cable to view the temperature and humidity change curves. GL100-WL can communicate wirelessly with a computer or Japanese Tuji GL240/GL840 through WIFI signals




  • Usage Introduction



  • ※1
    We use 2.4GHz radio waves.
    It can be used in China (excluding Hong Kong), Taiwan, Japan, the United States, Europe, and some regions.
    Please do not use it near the following environment, as communication speed may slow down due to radio interference and it may not be usable.
    ·Microwave oven
    ·Industry, Science, and Medical Equipment
    ·On the production line of the factory, a self established wireless telecommunications bureau (licensed wireless telecommunications bureau) is used to identify moving objects
    ·Specific small-scale power wireless telecommunications bureau (unlicensed wireless telecommunications bureau)
    ※2
    Condition:
    Temperature and humidity sensor (GS-TH) for recording and sampling temperature, humidity, dew point temperature, and accumulated temperature for 1 minute per file size, with a maximum file size of 1.9GB when connected to a power source
    ※3
    Please use the application software attached to the host to confirm the collected data.
    Condition 4: The CO2 sensor cannot use dry batteries.
    ※5
    Condition:
    Temperature and humidity sensor (GS-TH), 1-minute sampling, memory recording/no micro SD card/battery drive time varies depending on usage conditions and temperature
    Additionally, the battery drive time may vary depending on the module.





※1

Cannot connect multiple PCs to GL100 simultaneously

※2

Only GL100-WL supports LAN.

We use 2.4GHz radio waves.

Wireless LAN connection is only available in Japan, the United States, and Europe. Other regions cannot use it due to lack of recognition under the Radio Law.

Please do not use it near the following environment, as communication speed may slow down due to radio interference and it may not be usable.

·Microwave oven

·Industry, Science, and Medical Equipment

·On the production line of the factory, a self established wireless telecommunications bureau (licensed wireless telecommunications bureau) is used to identify moving objects

·Specific small-scale power wireless telecommunications bureau (unlicensed wireless telecommunications bureau)

※3

Having a fixed IP or DDNS server

Attention) When connecting to the DDNS server, it is necessary to set up transmission to GL100 on the router on the WAN side and LAN side.




  • TECHNICAL INDEX



GL100 host specifications and accessories

project content
Enter the number of ch According to different types of sensor modules
Input terminal shape
Input method
External input/output Alarm output 1ch
PC I/F USB (Micro B connector), wireless LAN (GL100-WL only) ON-OFF setting possible (factory set to ON)
storage device Built in RAM: approximately 4.9MByte
External external storage: Micro SD memory card
※ 1.9GB/data file
※ It is recommended to use the micro SD TS4GUSDHC4 (Transcend) that supports standby mode
Backup function Setting conditions: EEPROM
Timetable: Backup capacitor
Backup time for battery replacement: approximately 10 hours (at a temperature of 23 ℃)
monitor LCD (black and white with backlight, graphic dot matrix) (128 × 64)
show contents Measurement values, settings, etc
※ The collected data cannot be directly displayed on the host, please use the attachment.
wireless lan
(GL100-WL only)
Specifications: IEEE802.11b
Communication distance: about 40m
※ There may be differences depending on the usage environment. Real time transmission speed: up to 500ms
function USB I/F ·Read measurement data, host settings (using PC application)
·Set up network (GL100-Network_Config using PC application)
wireless lan ·Read measurement data, host settings (using PC application)
·Status monitoring, easy setup (using Android and iOS support for both end and end applications)
·Email sending and receiving function
Alarm output form Insulated switch type
Freeze frame: 30V, 50mA (allowable loss of 150mw)
trigger condition Start trigger: OFF · Level · Alarm · Date
Stop trigger: OFF · Level · Alarm · Date
judge Combination: Grade OR · Grade AND
Simulated CH judgment mode: uplink (↑) · downlink (↓)
Level trigger function 4VT, 4TSR, 3AT (edge action)
Store content Collect data
Easy waterproof host IP54
Only when the host is connected to the sensor module, the connector, and the battery cover are tightly closed.
Additionally, it is recommended to regularly replace the sealing strip
Shock resistance Equivalent to the first Class A of automotive components
power supply Dry battery 5, 2 batteries
USB bus power supply (5V 200mA or above)
When using an AC adapter for USB devices, a power supply of 5V.1A is required
AC adapter is not standard.
Scope of use environment -10-50 ℃, below 80% RH (non condensing)
Dimensions [W × D × H] 66mm × 100mm × 27mm (excluding protruding parts)
Weight (approximately) 125g (GL100-N), 130g (GL100-WL)
Accessories Instructions for use: 1 part
Guarantee letter: 1 unit
CD-ROM : 1 piece
USB (MicroB-A approximately 0.5m): 1 laptop

Temperature and humidity sensor GS-TH specification

project content
summary Temperature and humidity sensor GS-TH
A sensor module that can simultaneously measure temperature and humidity. Can simultaneously measure a wide range of temperature (-20~85 ℃) and humidity (0~100% RH)
Types of measurement Temperature, humidity, dew point temperature (calculated values)
Accumulated temperature ℃ h (calculated value)
Measurement method C-MOS sensor
Measuring temperature range -20℃~ 85℃
Temperature measurement accuracy -20 ≤ TS < 0 ± 0.8 (℃)
0 ≤ TS ≤ 60 ± 0.5 (℃)
60 < TS ≦ 85 ± 0.8 (℃)
Measure humidity range 0.0 ~ 100.0% RH
Measurement humidity accuracy 25℃
±10% RH (0 ~ 10% RH)
±8% RH (10 ~ 20% RH)
±5% RH (20 ~ 80% RH)
±8% RH (80 ~ 90% RH)
±10% RH (90 ~ 100% RH)
Response time Temperature and humidity: about 100 seconds
(63% responded with an air flow rate of 1m/s at 25 ℃)
※ There may be differences depending on the usage environment.
Calculation function Accumulated temperature: 0-99999k ℃ h
Reset calculation from inclusion (cumulative temperature exceeding or falling below the reference temperature)
sampling interval 0.5 · 1 · 2 · 5 · 10 · 20 · 30 seconds, 1 · 2 · 5 · 10 · 20 · 30 · 60 minutes
alarm OFF/Level
Temperature indicator ℃/°F
waterproof treatment The sensor area is covered with waterproof film
Usage environment Within the measuring range of the temperature and humidity meter (however, when placed in the same environment as the GL100 host, the host environment shall prevail)
Dimensions [W × D × H] 15mm × 45mm × 10.2mm (excluding protruding parts)
Weight (approximately) 14g

3-axis acceleration/temperature sensor GS-3AT specification

project content
outline 3-axis acceleration/temperature sensor GS-3AT
Sensor module capable of measuring 3-axis acceleration vibration (2G · 5G · 10G) and temperature (-10~50 ℃)
Types of measurement Acceleration and temperature of X · Y · Z 3-axis
Acceleration range ±2G(20m/s2)± 5G (50m/s)2)± 10G (100m/s)2)(Gravity acceleration 1G=approximately 9.807m/s)2
Accuracy of measuring acceleration
(including non-linear types)
±2% of F.S.
zone 0Hz~100Hz
interfere 25mG rms
Measuring temperature range -10 ℃~50 ℃<C-MOS sensor>
Temperature measurement accuracy -10 ≤ TS < 50 ± 1.0 (℃)
Response time (temperature) About 300 seconds
Measurement mode Storage mode, direct mode
storage capacity
(Storage Mode)
16 · 32 · 64 · 128k sampling
Pre trigger
(Storage Mode)
OFF/10·25·50·100%
sampling interval Storage mode 5·10 · 20 · 50 · 100 ms
direct mode 0.5 · 1 · 2 · 5 · 10 · 20 · 30 seconds, 1 · 2 · 5 · 10 · 20 · 30 · 60 minutes
Calculation function
(Direct mode)
performance Perform calculations on data obtained at 5ms intervals
Types of Algorithms 1-way calculation of average, maximum, and effective values (no calculation OFF)
Temperature indicator ℃/°F
OFFSETADJUST function ON ※ Function to adjust the current static acceleration value to 0
Dustproof and Waterproof IP54
Cable length About 20cm
Usage environment -10-50 ℃, below 80% (non condensing)
Dimensions [W × D × H] 51mm × 36mm × 19.5mm (excluding protruding parts)
Weight (approximately) 64g

4-ch voltage/temperature terminal GS-4VT specification

project content
summary 4-ch voltage/temperature terminal GS-4VT
It can perform temperature measurements with a maximum voltage of 4ch (20mV~50V and 1-5V) or thermocouple (K or T type). Each ch can choose voltage and temperature. In addition, logic (diagram) and pulse (counting) measurements can also be performed.
Types of measurement Voltage/temperature, logic/pulse (instantaneous, cumulative)
Measure ch Voltage/Temperature 4ch
Logic/pulse 4ch
Measure voltage range 20 · 50 · 100 · 200 · 500mV, 1 · 2 · 5 · 10 · 20 · 50V 1-5V F.S
Measurement voltage accuracy 0.15% of F.S.
Input impedance/input method 1M Ω± 5% full channel insulated input
temperature coefficient GAIN: 0.01% of F.S./℃ Zero: 0.02% of F.S./℃
Allowable signal source impedance Below 300 Ω
Maximum input voltage Input terminal+/- interval: 60Vp-p Input terminal/Input terminal interval: 60Vp-p Input terminal/GND interval: 60Vp-p
withstanding voltage Input terminal/between input terminals: 350Vp-p 1 minute Input terminal/GND: 350Vp-p 1 minute
Insulation resistance Between input terminal and GND: 50M Ω or above (DC500V)
COMMON mode removes ratio 90dB or above (50/60Hz signal source below 300 Ω)
interfere Above 48dB
Measuring temperature range <Thermocouple>K: -200.0 ℃~1370.0 ℃ T: - 200.0℃~ 400.0℃
Determination of temperature accuracy
(Reference contact compensation accuracy ± 0.5 ℃)
K-type thermocouple -200 ≤ TS ≤ -100 ± (0.05% of RDG+2.0 ℃)
-100 < TS ≤ 1370 ℃± (0.05% of RDG+1.0 ℃)
T-type thermocouple -200 ≤ TS ≤ -100 ± (0.1% of RDG+1.5 ℃)
-100 < TS ≤ 400 ℃± (0.1% of RDG+0.5 ℃)
Short circuit detection function have
Logic/pulse input Input Voltage 0~+24V (wire grounded input)
Input signal No voltage contact (contact a, contact b NO、NC)、 Open circuit collector, voltage input
Input threshold voltage Approximately+2.5V
Hysteresis value About 0.5V (+2V~+2.5V)
Pulse measurement range Instantaneous: Maximum 200C/sample
Cumulative value: Maximum 65535C (press QUIT key or start recording to reset)
Room temperature compensation ON / OFF
Temperature indicator ℃/°F
sampling interval 0.5 · 1 · 2 · 5 · 10 · 20 · 30 seconds, 1 · 2 · 5 · 10 · 20 · 30 · 60 minutes
Cable length About 20cm
Usage environment -10-50 ℃, below 80% RH (non condensing)
Dimensions [W × D × H] 46mm × 66mm × 35.5mm (excluding protruding parts)
Weight (approximately) 85g

4-ch temperature sensor input terminal GS-4TSR specification

project content
summary 4-ch temperature sensor input terminal GS-4TSR
Supports 2 types of thermistor sensors (optional) (GS-103AT-4P/GS-103JT-4P), capable of measuring temperatures up to 4ch (-40.0~105.0 ℃ (when GS-103AT-4P) or:- 40.0-120.0 ℃ (at GS-103JT-4P). In addition, logic (diagram) and pulse (counting) measurements can also be performed
Types of measurement Temperature, logic/pulse (instantaneous, cumulative)
Measure CH temperature 4ch
Logic/Pulse 4ch
Input method Scanning type, non insulated, 2 types of thermistor sensors
Measuring temperature range Supports 2 types of thermistor sensors (optional)
Type A:-40.0~105.0℃(GS-103AT-4P)
Type J:-40.0~120.0℃(GS-103JT-4P)
Measuring temperature Accuracy of the host part -40 ≤ TS < 0 ± 0.7 (℃)
0 ≤ TS ≤ 35 ± 0.2 (℃)
35 < TS ≦ 70 ± 0.4 (℃)
70 < TS ≦ 120 ± 1.0 (℃)
thermistor ± 0.4 ℃ (representative value)
GS-103AT-4P / GS-103JT-4P
Logic/pulse input Input Voltage 0~+24V (wire grounded input)
Input signal No voltage contact (contact a, contact b NO、NC)、 Open circuit collector, voltage input
Input threshold voltage Approximately+2.5V
Hysteresis value About 0.5V (+2V~+2.5V)
Pulse measurement range Instantaneous: Maximum 200C/sample
Accumulated value: Maximum 65535C (can be cleared by pressing the 【 QUIT 】 key or starting recording)
Temperature indicator ℃/°F
sampling interval 0.5 · 1 · 2 · 5 · 10 · 20 · 30 seconds, 1 · 2 · 5 · 10 · 20 · 30 · 60 minutes
Cable length About 20cm
Usage environment -10-50 ℃, below 80% RH (non condensing)
Dimensions [W × D × H] 46mm × 66mm × 35.5mm (excluding protruding parts)
Weight (approximately) 83g

Illumination/UV Sensor GS-LXUV Specification

project content
summary Illumination/UV sensor GS-LXUV
Can simultaneously measure illuminance (0 lx~200k lx) and ultraviolet radiation (0~30mW/cm)2)The sensor.
Types of measurement Illumination, UV intensity, cumulative illumination (cumulative display only in inclusion), cumulative UV intensity (cumulative display only in inclusion)
Measure ch illuminance 1ch
UV intensity 1ch
measuring range illuminance 0 lx ~ 200k lx
UV intensity 0 ~ 30mW/cm2
measurement accuracy illuminance ±5%
UV intensity ±5%
Spectral accuracy illuminance 400-750nm (up to 560nm)
UV intensity 260-400nm (up to 370nm)
Determination of decomposition energy illuminance 0.1 lx
UV intensity 0.001mW/cm2
responsiveness illuminance Less than 1ms
UV intensity Less than 1ms
Calculation function Accumulated illuminance 0 ~ 999,999klxh
Accumulated UV intensity 0 ~ 999,999W/cm2h
※ You can re count by pressing the [QUIT] key or starting recording
sampling interval 0.5 · 1 · 2 · 5 · 10 · 20 · 30 seconds, 1 · 2 · 5 · 10 · 20 · 30 · 60 minutes
Cable length About 20cm
Usage environment -10-50 ℃, below 80% RH (non condensing)
Dimensions [W × D × H] 46mm × 66mm × 19mm (excluding protruding parts)
Weight (approximately) 65g

CO2 Sensor GS-CO2 Specification

project content
outline CO2 sensor GS-CO2
Can measure CO2Sensors within the concentration range of 0-9999ppm.
Sensor method NDIR method (non dispersive infrared absorption method)
Types of measurement CO2concentration
Measure ch 1ch
Measurement concentration range 0~9999ppm (5001~9999ppm is outside of ensuring accuracy)
Measurement of concentration accuracy ±(5% of rdg + 30ppm)0~5000ppm
Responsiveness Less than 3 points
Sampling interval 0.5 · 1 · 2 · 5 · 10 · 20 · 30 seconds
1 · 2 · 5 · 10 · 20 · 30 · 60 min (GS-CO2 sampling is 2 sec)
calibration ON / OFF
The length of the wire About 20cm
Usage environment 0-50 ℃, below 80% RH (non condensing)
Dimensions [W × D × H] 66mm × 100mm × 24.6mm (excluding protruding parts)
Weight (approximately) 99g

AC AC current sensor adapter GS-DPA-AC specification

project content
outline AC AC current sensor adapter GS-DPA-AC
The AC current sensor can measure the power of single-phase 2-wire, single-phase 3-wire, three-phase 3-wire, etc. by connecting the 50A clamp current probe (GS-AC50A), 100A clamp current probe (GS-AC100A), and 200A clamp current probe (GS-AC200A).
Connectable sensors 50A clamp type current probe (GS-AC50A)
100A clamp type current probe (GS-AC100A)
200A clamp type current probe (GS-AC200A)
※ Options
Types of measurement Current, electricity, cumulative electricity (cumulative display is only available for inclusion)
Measure ch The maximum number of sensors that can be connected is 2ch
Input method RMS converter+scanning, non insulated input 3 types of AC sensors connected
Measurement accuracy of current <Body+Sensor>
±2.0% FS ± 1digit
Surrounding temperature of 23 ℃, freeze input, freeze frequency
Measure the frequency of the object being measured 50Hz / 60Hz
Applicable circuit Single phase 2-wire single-phase 3-wire or three-phase 3-wire
Measurement value display function Current, electricity (instantaneous electricity, cumulative electricity)
Calculation function Accumulated electricity: 0~999999 kWh * recalculated from inclusion onwards
Measure grid current once GS-AC50A :50A
GS-AC100A:100A
GS-AC200A:200A
Voltage endurance AC1000V、50/60Hz、1min
Between the body and the input/output terminals
Sampling interval 0.5 · 1 · 2 · 5 · 10 · 20 · 30 seconds 1 · 2 · 5 · 10 · 20 · 30 · 60 min
Cable length Adapter for AC current sensor About 20cm
AC current sensor About 20cm
Usage environment -10-50 ℃, below 80% RH (non condensing)
Dimensions [W × D × H] 46mm × 66mm × 27.4mm (excluding protruding parts)
Weight (approximately) 55g

GS AC current sensor adapter GS-DPA specification

project content
summary GS bifurcation adapter GS-DPA
A converter that can freely combine two out of three temperature and humidity sensors (GS-TH), CO2 sensors (GS-CO2), and illuminance/ultraviolet sensors (GS-LXUV) for measurement.
Number of connectors 2ch
Sensors that can be connected GS-TH, GS-CO2, GS-LXUV (cannot use the same sensor)
Cable length About 20cm
Usage environment -10-50 ℃, below 80% RH (non condensing)
Dimensions [W × D × H] 46mm × 66mm × 27.4mm (excluding protruding parts)
Weight (approximately) 68g

GL100-APS specification

project content
Support OS Windows 8.1/8/7/Vista (64/32bit)
Connection method USB、 wireless lan
function Host control, real-time data recording, file conversion, data regeneration
Ontology setting scope Input settings, recording settings, alarm settings, trigger settings
Collect data Real time data (GBD binary CSV)
Host records data
SD card data
show contents Analog waveform, logic waveform, pulse waveform, digital value
display mode Y-T display, numerical display, statistical/historical display, X-Y display between cursor (only for playback)
file conversion Cursor spacing, full data, playback function
monitoring function Alert monitoring regularly sends messages to designated email addresses.
Report function Automatically generate daily and monthly report files

GL100 Network Config Specification

project content
Support OS Windows 8.1/8/7/Vista (64/32bit)
Connection method USB
function USB driver installation, GL100 update and upgrade
Time synchronization function, machine name setting, wireless LAN setting, email sending and receiving setting

Software specifications for smart devices

project content
Support OS iOS(7.0~)、Android(4.3~)
Connection method wireless lan
function Digital display, waveform display (only for playback), recording start/stop control, sampling interval update, email transmission control function, customization function



Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!