Karl Fischer method moisture analyzer [volumetric titration method] MKV-710D

Support data integrity, including access restrictions, permission management, audit trails, electronic records/signatures, backup/restore/archive, data capacity, security settings, LIMS compatibility, and other functions.
Using a large color touch screen with wireless connection, it can control two measuring units.
MKV-710M/S can be upgraded to the MKV-710D model through upgrade settings.

Karl Fischer method moisture analyzer [volumetric titration method] MKV-710D main features:

1. Adopting a large 8.4-inch color LCD touch screen for operation and control.
2. The touch screen is more secure and reduces the risk through wireless Bluetooth.
3. The new burette unit reduces dead volume and reagent usage.
4. Control and charge the second set of driving devices, which can automatically perform titration calibration.
5. The measurement results can be stored on a USB drive and generate a PDF experimental report.
6. Real time display of moisture values and curves. Choose one or two sets of burettes.
7. Different solvents can be replaced without changing the polarization current or endpoint settings.

Karl Fischer method moisture analyzer [volumetric titration method] MKV-710D technical parameters:

Name and Model

Karl Fischer Moisture Analyzer [Volumetric Titration Method] MKV-710D

measuring range

water content: 0.1~500mgH2O

Moisture content: 1ppm~100%H2O

Accuracy of burette

Volume of burette: 10mL, accuracy: ± 0.015mL, repeatability: ±0.005mL

Endpoint detection

Polarization potential detection using dual platinum electrodes

Terminal judgment

The duration of the polarized potential within the specified range is determined as the endpoint

Titration method

Positive titration/reverse titration (requires additional purchase of a second set of burettes)

Solvent demand

30mL~100mL (using an S-shaped titration cup)

Other functions

1) Drift value automatic adjustment

2) Automatic start measurement

3) Can store 10 sets of blank values

4) Optional automatic or manual suction and discharge device

Key operation

Touch screen direct operation

display

1) 8.4-inch color LCD screen, resolution: 800×600

2) Chinese/English/Japanese operating interface

3) Dual channel simultaneous display

calculation

Concentration calculation, statistical calculation, automatic input of blank value and titration degree

Data storage

120 measurement methods, 5000 sample results, 100000 operation records

Data integrity function

Access restriction: Authentication through personnel ID and password

Permission management: Assign each operator to a specific permission group

Audit Tracking: Automatically Record Historical Operation Information

Electronic Records/Electronic Signature: Electronic approval and signature based on workflow

Backup/Restore/Archive: Automatically backup data at any interval

External output/input

RS-232C×4,USB×1,SS-BUS×1,LAN×1