LI-7500 series (including LI-7500, LI-7500A, LI-7500RS, LI-7500DS) open circuit CO2/H2The O gas analyzer has been widely used in the field of global vorticity covariance carbon flux monitoring. More than 90% of the flux observation stations of the Global Carbon Flux Observation Network (FLUXNET) use the LI-7500 series CO2/H2O gas analyzer.
LI-7500DS is a new generation of digital open circuit CO2/H2The O intelligent analysis system is designed for rapid measurement of CO in the atmospheric environment2And H2Designed by O,The low power consumption of up to 4W and almost maintenance free characteristics make it more suitable for long-term stable measurements at field observation stations.
main features
|
Low power consumption:Ensure that it can perform long-term positioning measurements on field stations |
|
Continuous measurement in all directions:It can sample and measure airflow in all directions, allowing for multiple installation options and ensuring continuous data measurement |
|
Record complete vorticity covariance dataIncluding wind speed data collected by ultrasonic anemometers and auxiliary meteorological, radiation, and soil data collected by LI-COR Biomet system |
built-inSmartFlux®system
SmartFlux®The system will integrate EddyPro®The functionality of the software is brought to your field site, like a powerful field computer that can perform real-time and complete vorticity flux calculations. The calculation results are consistent with your use of EddyPro in the office®The results obtained from processing raw data are exactly the same, but more timely. SmartFlux®The system is FluxSuite®The hardware foundation of the data online monitoring and management system.
TECHNICAL INDEX
CO2measure
·Calibration range: 0~3000 µ mol/mol
·Accuracy:< 1%
·Zero drift (per ℃): Typical ± 0.1 µ mol/mol, maximum ± 0.3 µ mol/mol
·RMS noise/resolution (typical @ 370 µ mol/mol CO)2):
o @5 Hz:0.08µmol/mol
o @10 Hz:0.11µmol/mol
o @20 Hz:0.16µmol/mol
·Gain drift (% of reading per ℃ @ 370 µ mol/mol):
Typical ± 0.02%
Maximum ± 0.1%
·Regarding H2Sensitivity of O (mol CO)2/mol H2O):
Typical ± 2.00E-05
Maximum ± 4.00E-05
H2Omeasure
·Calibration range: 0~60 mmol/mol
·Accuracy:< 1%
·Zero drift (per ℃):
Typical ± 0.03 mmol/mol
Maximum ± 0.05 mmol/mol
·RMS noise/resolution (typical @ 10mmol/molH)2O):
o @5 Hz:0.0034 mmol/mol
o @10 Hz:0.0047 mmol/mol
o @20 Hz:0.0067 mmol/mol
·Gain drift (% of reading per ℃ @ 20 mmol/mol):
Typical ± 0.15%
Maximum ± 0.30%
·Regarding CO2Sensitivity (mol H)2O/mol CO2):
Typical ± 0.02
Maximum ± 0.05
comprehensive
·Basic gas sampling frequency: 150Hz. Bandwidth: 5,10 or 20Hz;
·Analyzer: Absolute Open Circuit Non Dispersive Infrared Solid State Detector
·Detector: Thermoelectric cooled lead selenide optical path length 12.5cm
·Data communication: Ethernet
·Air temperature thermistor: 10K Ω @ 25 ℃ thermistor
Measurement range:- 40~70℃
Accuracy: ± 0.25 ℃ @ -20~70 ℃
Resolution: 0.003 ℃ @ 25 ℃
·Pressure sensor
Measurement range: 20-110 kPa
Accuracy: ± 0.4 kPa @ 50~110 kPa
Resolution: 0.006kPa
·Working temperature range: -25 to 50 ℃ (can be expanded to -40~50 ℃ as needed)
·Relative humidity range: 0-95% (non condensing)
·Waterproof rating IEC IP65
·User Interface: Windows Environment PC Software
·Power supply: 10.5~30 VDC
·Power consumption: Typical 4W @ 25 ℃, maximum 8W @ -25 to 50 ° C during operation
Analyzer probe
·Size: Diameter 6.5 cm, Length 30 cm
·Weight: 0.67 kg; Connect the base 1.3kg; Cable length 200cm
DSIbox
·Size: 13.24cm x 14.64cm x 6.24cm (H x W x D)
·Weight: 0.93kg
Origin and Manufacturer: United StatesLI-CORcompany