LNG gas analyzer
Execution standard for LNG gas analyzer:
Product details

LNG gas analyzer

LNG gas analyzer1 cubic meter of natural gas calorific value is an instrument that can be used to quickly calculate characteristic indices such as high calorific value, low calorific value, density, relative density, Huabai number, combustion potential, etc. of different gases. It is a manufacturer of natural gas component content analyzers and can be widely used in gas appliance production enterprises, gas measurement and testing departments, gas filling stations, scientific research, environmental protection, and gas distribution industries.

LNG gas analyzerExecution standards:

The instrument for measuring and analyzing methane and nitrogen content in natural gas is configured to meet national standards
1. The natural gas analyzer complies with GB/T13610-2003 "Composition Analysis of Natural Gas - Gas Chromatography Method"
2. The instrument complies with SH/T0230-92 "Determination of Composition of Liquefied Petroleum Gas (Chromatography Method)

Requirements for laboratory installation conditions:

Power supply voltage: 220V~± 22V, 50Hz ± 0.5Hz

Total power of the instrument: ≤ 2000W

Environmental temperature:+ 5℃~+35℃

Relative humidity:< 85%

Attention: There should be no corrosive gases or electric or magnetic fields that affect the normal operation of the instrument in the placement location. The instrument placement workbench should be stable and free from vibration.

Precautions for:

1. Standard gas should be used within its validity period.

2. The time interval for instrument calibration should be determined based on the level of the measuring station and the performance of the instrument.

3. When replacing the carrier gas cylinder, attention should be paid to avoiding air from entering the gas system to prevent damage to the detector of the instrument.

4. When the temperature is very low in winter, the heating and insulation device of the sampling system should be activated in a timely manner to prevent sample condensation.

5. Regularly check the filter element of the filter and replace it if necessary.