Product features:
1. Independent negative pressure circulation structures are installed throughout the workspace and transfer window to maximize the protection of personnel safety
2. Unlike traditional secondary biosafety cabinets, the exhaust gas from the tertiary cabinet is filtered through two ULPA ultra high efficiency filters and then discharged outdoors to achieve complete filtration of biological hazardous particles, providing dual protection for personnel and the environment
3. The interior and exterior of the cabinet can withstand H2O2 sterilization
4. Full external discharge structure, capable of handling toxic samples and volatile radioactive nuclides in biological experiments
5. German EBM company RadiCal series centrifugal external rotor fan, low noise, low maintenance cost
6. The ultra large screen microcomputer program control system can display real-time operating parameters inside the cabinet, display the working area in real time, transmit window pressure and settling airflow velocity, and provide sound and light alarms when there is excessive fluctuation.
7. A pre filter is installed on the top of the cabinet to prevent large dust particles from clogging the main filter in the incoming airflow, effectively extending the service life of the main filter
8. The whole machine is equipped with four Swedish Camfil ULPA ultra high efficiency air filters, which effectively increase the interception area and extend its service life
9. Provide vertical layer airflow to the workbench to protect samples. Both the operating room and transfer chamber can meet ISO 14644.1 international standard level 3, which is 100 times cleaner than level 100
10. Provide chlorosulfonated polyethylene rubber sleeves and nitrile gloves in different sizes, easy to replace
11. IsocideTMThe surface antibacterial coating can effectively inhibit the growth of microorganisms on the surface of the cabinet
model |
External dimensions (mm) (Width x Depth x Height) |
Internal dimensions (mm) (Width x Depth x Height) |
Internal dimensions of transfer window (mm) (Width x Depth x Height) |
Settlement wind speed |
Remove a single glove The intake wind speed |
SCI-2G_-_ | 1645 x 845 x 2273 | 965 x 620 x 620 | 433x620x620 | 0.4m/s | ≥0.7m/s |
SCI-3G_-_ | 1950 x 845 x 2273 | 1270x 620 x 620 | 433x620x620 | 0.4m/s | ≥0.7m/s |