Qixiang Tengda (Beijing) Technology Co., Ltd
Home>Products>MCW oil spectrometer
Product Groups
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
  • Address
Contact Now
MCW oil spectrometer
MCW oil spectrometer
Product details

Instrument features:

1) completely satisfiedASTM D6595andNB/SH/T 0865-2013Standard testing methods

2) Obtained the US military standardJOAPCertification (US Department of Defense Equipment Number:6650015660884

3) No need for traditional tracing and electrode compensation operations.

4) Standard configuration can be tested simultaneously24Various analytical elements can be flexibly selected and added according to different needs, up to a maximum of simultaneous measurements31Seed elements

5) Thermostatic control of optical system:40°C±1°C

6) A position sensor is installed in the sample burning room to measure the position of the oil cup, the gap between the placement points, and the position of the electrode on the disk rod

7) No need for dedicated grounding device

8) High resolution dual layer15sliceCCDSensors, adding elements without hardware changes

9) The equipment has an exhaust structure

10) Low oil sample consumption:<1.5ml

11) Standard configurationASTMandSHStandard requirements375pxLong high-purity graphite rod electrode and disc electrode

12) Integrated design, built-in industrial touchscreen computer and dedicated printer

13) Easy to operate, operators do not require professional training

14) No need for sample pre-treatment

15) Short detection time: analysis time less than40second

16) We can provide specialized testing procedures and standard substances for other industrial fluids such as antifreeze, industrial circulating water, fuel oil, etc

17) No need for auxiliary working gas, can be used in general office environments

Sample Room

1) Graphite disk electrode,modelASTM D-2diameter0.492″×wide0.2″(ASTMStandard)

2) Graphite rod electrode,diameter0.243″×long6″ASTMStandard)

3) Rod shaped electrode holder with automatic adjustment of electrode distance device

4) Activate the safety switch of the room door

5) Oil collection disc

6) Visual window, visually observing the entire excitation process

Excitation light source

1) According to the US military standardsJOAPDefined oscillating arc discharge

2) Equipped with a stable voltage device

3) Solid state source excitation ensures constant current source frequency output without the need for periodic cleaning or cleaning of excitation source errors

4) Equipped with internally adjustable functions

Control system software

1) Pre installedWindows7Operating System (English Version)

2) Collect all spectral dropdown menus and combine themwindowWindow display

3) Data output format: AETCSIDCSVGenericWaiting for output format

4) The content and format of the saved results can be set

5) It can directly display various raw data such as optical density and concentration of each element

6) Can automatically recalculate results, including mean, standard deviation, and relative standard deviation

Power requirements

1) Power supply:240 VAC+/- 10%50 Hertz

2) Consumption during sample analysis work500Below the tile

3) Instrument consumption while waiting for use275Below the tile

4) Built in power circuit breaker protection device and switch

Environmental requirements

1) Environmental temperature:530

2) Relative humidity:≤90%

optical system

1) usePaschen-RungeOptical system, high precision and stability

2) Thermostatic control of optical system:40°C±1°C

3) Spectral range:203~810nm

4) Temperature stable materials and structural design

5) High resolution dual layer15sliceCCDsensor

6) Without adding any hardware, adding software and standard oil can expand to31Ability to test elements

7) Equipped with dynamic background correction and automatic peak finding function

Instrument specifications

1) Size: Instrument77.5×63.5×70 cm

2) Net weight:116 kg

3) Gross weight:185 kg

Detecting elements and scope (standard configuration)

Serial number

element

Detection range, ppm

1

Al

aluminium

0-1,000

2

C

carbon

Ref.

3

Ba

barium

5-6,000

4

B

boron

0-1,000

5

Ca

calcium

0-6,000

6

Cd

cadmium

0-1,000

7

Cr

chromium

0-1,000

8

Cu

copper

0-1,000

9

H

hydrogen

Ref.

10

Fe

iron

0-1,000

11

Pb

lead

0-1,000

12

Mg

magnesium

0-6,000

13

Mn

manganese

0-1,000

14

Mo

molybdenum

0-1,000

15

Ni

nickel

0-1,000

16

P

phosphorus

5-6,000

17

Si

silicon

0-1,000

18

Ag

silver

0-500

19

Na

sodium

0-6,000

20

Sn

tin

0-1,000

21

Ti

titanium

0-1,000

22

V

vanadium

0-1,000

23

Zn

zinc

0-6,000

24

K

potassium

0-1,000

25

Li

lithium

0-1,000

26

Sb

antimony

0-1,000