MD8H Practical Microwave Digestion Instrumenttechnical parameter


1. Dual measurement and control of temperature and pressure;

2. High precision high-temperature melt pressure sensor, with a pressure measurement range of 0-10MPa (1500Psi) and a pressure measurement accuracy of ± 0.01MPa;

3. High sensitivity platinum resistance temperature sensor, temperature measurement range 0-300 ℃, monitoring accuracy ± 0.1 ℃;

4. Dissolution tank, vertical explosion-proof design, easy disassembly and assembly, fast cooling; Double pressure protection digestion tank with vent hole and sealed bowl. Capacity of 100mL, pressure resistance of 10Mpa, temperature resistance of 300 ℃.

5. Configure to process 1-8 samples simultaneously, with 5 layers of Teflon coating in the inner cavity for ultra strong acid corrosion resistance, ensuring at least 5 years of corrosion resistance;

6. Variable frequency circuit, outputting continuous microwave, maximum power of 1600W, effective power of 1200W, resonant microwave and power matching microwave cavity design to ensure microwave uniformity;

7. The furnace chamber ventilation adopts acid resistant and high air volume centrifugal fans with an exhaust volume of 5m3/min;

8. Specially designed self-locking safety sliding door, capable of withstanding accidental explosions.




MD8H Practical Microwave Digestion Instrumentmain features


1. Adopting a large cavity design that is different from household microwave ovens and has completely independent intellectual property rights, it can simultaneously place 1-8 100ml ultra-high pressure digestion tanks;

2. Adopting Japanese Panasonic magnetron (5-year warranty) and professional uniform cavity (microwave resonant cavity); Industrial grade 1600W magnetron ensures long-term stable operation of the instrument;

3. Dual temperature and pressure measurement and control, precise control, actively preventing overheating and overpressure

4. High precision high-temperature melt pressure sensor, using ideal gas control method, directly measures the actual reaction pressure inside the digestion tank, with a pressure measurement range of 0-10MPa and a pressure measurement accuracy of 0.01MPa;

5. Plug in high-sensitivity platinum resistance temperature sensor, directly measuring the actual reaction temperature inside the digestion tank, with a temperature measurement range of 0-300 ℃ and a temperature measurement accuracy of ± 0.1 ℃;

6. It can measure and control the pressure and temperature inside any reaction tank, accurately displaying the chemical reaction process;

7. Multiple materials and different capacities can be selected, and a perfluorinated digestion tank or an FT double-layer digestion tank can be configured: the inner tank of the FT digestion tank is made of PTFE (polytetrafluoroethylene), and the sheath is made of imported material PEEK one-time die-casting, which is resistant to high temperature, high pressure, and strong acid corrosion; Easy disassembly and assembly, fast cooling; Double pressure protection digestion tank with vent hole and sealed bowl; Up to 8 ultra-high pressure digestion tanks can be placed simultaneously at once;