Taizhou Meixu Instrument Equipment Co., Ltd
Home>Products>MX-2000S UV Fluorescence Sulfur Analyzer
MX-2000S UV Fluorescence Sulfur Analyzer
UV Fluorescence Sulfur Analyzer SH/T 0689 ASTM D5453,GB/T 17010, Widely used for detecting sulfur content in liquid, solid, or gas samples.
Product details

undefined
undefined
undefined
undefined
undefined
undefined

Product Overview:

MX-2000S UV Fluorescence Sulfur Analyzer SH/T 0689 ASTM D5453,GB/T 17010

Domestic integrated sulfur analysis instruments are widely used for detecting sulfur content in liquid, solid, or gas samples. Compared with similar instruments at home and abroad

It has the characteristics of stable and reliable performance, high analysis accuracy, and good repeatability.
The total sulfur content in the sample was rapidly determined by UV fluorescence method. The application of high-precision data acquisition cards and computer technology for data acquisition

Collecting, controlling, and processing provide reliable guarantees.

The instrument system uses ultraviolet fluorescence method to determine the total sulfur content, which improves the ability of Kang impurities to interfere and avoids the cumbersome operation and potential issues of the titration cell caused by the electric method

The instability caused by this greatly improves the sensitivity of the instrument. The key components of the system use Jin port devices, which provide reliable overall performance

Guarantee.


Instrument execution standard:

SH/T 0689, ASTM D5453 Determination of Total Sulfur Content in Light Hydrocarbons, Engine Fuels, and Other Oil Products (Ultraviolet Fluorescence Method)

Technical Parameter

Basic parameters: Sample types: liquid, solid, and gas
Measurement method: UV fluorescence method (S)
Sample injection volume: Solid sample: 0.1-5mg
Liquid sample: 1-20 μ L
Gas sample: 1-5mL
Measurement range: S:0.2 ~10000mg/L
Measurement accuracy:

Concentration value (ppm)

Injection volume (μ L)

RSD(%)

0.2

20

25

5

10

10

50

10

5

100

10

3

5000

10

3

Temperature control range: Room temperature~1050 ℃
Temperature control accuracy: ± 3 ℃

Sample injection volume: Liquid: less than 20uL

Gas source requirements: High purity argon gas with a purity of over 99.995%,
High purity oxygen: purity above 99.995%
Power supply: AC220V ± 22V, 50Hz ± 0.5Hz, 1500 W

Measurable sample status: liquid, gas (with corresponding injector), PMT high pressure range: DC500V~900V, According to the measured concentration, pass

Set the required values for the operating system.

Requirements for instrument completeness: host (data acquisition, control system), cracking furnace, automatic sampler (gas, liquid), computer (including operating software)

Wait.

Dimensions: Host: 305 (W) × 460 (D) × 440 (H) mm
Temperature control: 550 (W) × 460 (D) × 440 (H) mm
Weight: Host: 20kg
Temperature control: 40kg

The instrument is suitable for measuring the total sulfur of paraffin oil, diesel, gasoline, lubricating oil, fuel oil, liquefied gas, as well as other oil products, chemical raw materials, and finished products

Content.

undefined
undefined

undefined

Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!