Hebi Lixin Instrument Co., Ltd
Home>Products>Microcomputer based glial layer analyzer
Microcomputer based glial layer analyzer
A specialized instrument used for measuring the gelatinous layer index of bituminous coal, including three indicators: the maximum thickness Y of the
Product details

Microcomputer based glial layer analyzer


Scope of application:

A specialized instrument used for measuring the gelatinous layer index of bituminous coal, including three indicators: the maximum thickness Y of the gelatinous layer, the maximum final shrinkage X, and the type of volume curve. The equipped controller consists of two separate systems, which can control two sets of heating furnaces simultaneously or one set of heating furnaces separately. Widely used in coal, metallurgy, chemical, geological exploration and other departments.

Performance characteristics:

1. The bidirectional communication between the computer and the controller enables comprehensive monitoring of the control equipment.

2. Automatically control and display the heating temperature of single or double furnaces.

3. Automatically measure and display the shrinkage of coal samples from single or double furnaces.

4. Automatically draw volume curves for single or double furnaces.

5. Automatically draw the curve of furnace temperature variation over time.

6. Automatically fill in the furnace temperature before and after every ten minutes.

7. Alarm at the inflection point of the volume curve, reminding laboratory personnel to perform operations such as starting, stopping, silencing, and running the equipment.

8. Manually detect the data of the adhesive surface, input it into the computer through the keyboard and save it.

9. Automatically draw upper and lower adhesive surface curves and calculate the maximum thickness of the adhesive surface.

10. Automatically calculate test results and print out test records and reports.

11. Store previous test data for querying and printing.

TECHNICAL INDEX

Temperature measurement range: 0-999 ℃, resolution ± 1 ℃, equipped with K-value thermocouple

Time measurement range: 0-999min, resolution 0.1min

Timing error: ± 30s/24h

Quartz recording drum linear speed: 1mm/min

Power: ≤ 5KW

Controller accuracy: Level 0.5

Accuracy: Draw a length of 160 ± 2mm in 160 minutes

Temperature resolution: 1 ℃

Heating control: reach 250 ℃ within 30 minutes, 3 ℃/min between 250 ℃ and 730 ℃