Microcoulometric chlorine content analyzer
Product Description:
The instrument adopts the principle of microcoulometry and is used for the determination of total chlorine content in products such as petrochemicals, coal chemicals, aromatic hydrocarbons, and natural gas. It has the advantages of high sensitivity and good repeatability.
Applicable standards:
Chlorine: SH/T1757、ASTM D5808
Main configuration:Microcoulometric titrator mainframe, stirrer, sampler, computer, etc
Technical parameters:
1. Sample status: gas, liquid, solid (with corresponding injector optional)
2. Gas source requirements: nitrogen purity of 99.95%, output pressure ≤ 0.2MPa
Oxygen purity 99.95%, output pressure ≤ 0.2MPa
3. Analysis time: 2 minutes
4. Support LIMS system connection
5. Power supply: AC 220V ± 22V, 50Hz ± 0.5Hz; Power: 3000W
6. Measurement range:Cl: 0.3mg/L~10000mg/L~percentage content
7. Repetitive error:
0.1mg/L ≤ sample (or standard) concentration < 1.0 mg/L, ≤ ±0.1 mg/L
1.0mg/L ≤ sample (or standard) concentration ≤ 10 mg/L, ≤ 10%
Sample (or standard) concentration>10 mg/L, ≤ 5%
Performance characteristics:
1. The quartz cracking tube adopts imported high-purity quartz material to improve the accuracy and repeatability of detection.
3. Adopting corrosion-resistant electrode caps to avoid the problem of large detection result errors caused by electrolyte corrosion of electrode caps and poor electrode contact.
4. The chlorine content detection of the microcoulometric titrator adopts a one-way valve dehydration device to remove moisture from the sample gas and improve the accuracy of sample detection.