Product Introduction
Milli-Q ™ Advantage A10 ™ The system uses pure water reverse osmosis water, distilled water, deionized water, or Elix pre treated water as the inlet source to produce high-quality first grade ultrapure water (with a resistivity value of 18.2 M Ω)· cm@25 ℃, TOC 5 ppb), The maximum flow rate can reach 2L/min. Two independent units - Milli-Q ™ Advantage A10 ™ Water production host and Q-POD ™ The water dispenser can provide ultra pure water solutions suitable for different laboratory professional applications.
main features
1. Stable and efficient production of ultrapure water
1) Pure water from Elix ™ Pure water system is preferred to enter Q-Gard according to the selected water source ™ The initial purification column and the pre purified water flow are then passed through a dual wavelength ultraviolet lamp to ensure that organic molecules are oxidized and bacteria are removed. Continuing with Quantum ™ The purification column removes ions and organic pollutants below trace levels to provide ultra pure water quality that meets application requirements. Ultra pure water meets the first level water quality standards set by ASTM, CAP, NCCL, EP, and USP.
2) Multiple water qualities can be combined with a series of terminal refiners to meet the needs of different specific applications
2. Comfortable and flexible water intake
1) Q-POD water dispenser provides convenient and fast access to Class I ultrapure water
2) The POD water dispenser occupies a small experimental platform area or can also be installed on the wall for easy access to pure water anytime and anywhere
3) Up to 3 POD water dispensers can be connected at various locations in the laboratory
4) The POD water intake handle can rotate 360 ° and adjust its height up and down, suitable for all conventional laboratory vessels for water intake
5) The color display screen of the POD water dispenser keeps you informed of key water quality information to ensure optimal water quality
6) Accurate flow meter
Manual water intake can be achieved by pressing the switch on the Q-POD water dispenser to obtain pure water from drop by drop to high flow rate. The water intake arm can also be removed from the bracket to facilitate your application and meet various laboratory needs
The control panel of Q-POD is equipped with a quantitative water intake button. Simply press the+or - button and then press the quantitative water intake button to accurately and accurately set the quantitative water intake. It can be reused once
3. Dual water quality monitoring
The comprehensive water quality monitoring patented TOC detector and high-precision resistivity detector ensure stable and accurate water quality monitoring
1) Accurate and effective organic matter monitoring
The TOC detector uses a 0.5 ml quartz tube to obtain ultrapure water. A10 ™ When the ultraviolet lamp is turned on, organic compounds undergo photocatalytic oxidation reactions. The final product of organic matter oxidation is the increase in conductivity caused by the dissolution of carbon dioxide in water. The titanium electrode in the TOC detector can continuously monitor the temperature compensation of the conductivity change up to 25 ℃ through a series of precise algorithms to ensure complete oxidation and calculate the carbon content related to the conductivity change.
A10 ™ The patented technology of TOC detector has the following advantages
The oxidation process and conductivity detection are carried out in the same quartz tube. Finally A10 ™ The detector will confirm that all organic compounds have been oxidized to terminate the reaction and provide accurate and reproducible TOC values.
The externally calibrated TOC detector with a certificate has a detection range of 1~999 ppb.
• Design to allow TOC adaptability testing to be performed in accordance with USP § 643 and EP European Pharmacopoeia requirements
2) High precision ion detection
Milli-Q ™ The high-precision resistivity detector of Integral system has the following specific functions and advanced design to ensure accuracy
Ensure the accuracy of the displayed values on the system screen.
The patented design of coaxial electrodes ensures the stability of electrode constants.
The design of the flow formula without dead corners ensures that the detection results can accurately reflect the actual ion concentration in the water.
Low electrode constant (0.01 cm-1) ensures optimal detection accuracy under low ion contamination in accordance with ASTM ® Requirement of D 1125-95 (2009).
Temperature sensitivity up to 0.1 ℃ can accurately display temperature compensated resistivity ASTM ® D 1125-95
(2009) Standard recommendation.
Automatic alarm when resistivity detection is abnormal
The design allows for resistance adaptability testing to be performed in accordance with USP § 645 and EP European Pharmacopoeia requirements.
3) RFID tag technology
Ensure the safety and traceability of the purification column
4. Intelligent system management
Provide three levels of information management
1) All necessary information for regular use is directly displayed on the screen of the POD water dispenser
2) The system maintenance main screen will display prompt information for each step to guide you on how to proceed to the next operation
3) System management protects key parameters such as set points through the ID and password settings in the "Administrator" menu
5. Intelligent system maintenance
1) The unique 'laboratory shutdown mode' allows you to rest assured even during long holidays
2) Built in computer chip can record water quality reports for up to 2 years
3) Automatic alarm and prompt to promptly remind users to perform system maintenance
4) Can be connected to an external printer for real-time printing of water quality reports
Application scope
Revolutionary Q-POD ™ Technology provides convenient and flexible water intake options. Each Milli-Q ™ Advantage A10 ™ The system can place up to three Q-PODs in different locations within the same laboratory ™ Device. Different applications may focus on specific pollutants in water. Milli-Q ™ Advantage A10 ™ The system provides a unique series of terminal refiners, which can be selected according to the application to remove specific pollutants that may affect the results
1. Selecting a 0.2 micron terminal filter to produce water is suitable for most physicochemical analysis applications such as HPLC, LC-MS, MALDI-TOF-MS, IC, ICP, AA, etc
2. BioPak water can be used for molecular biology applications and cell culture
3. Experimental study on EDS endocrine disruptors using EDS Pak produced water for environmental analysis
4. Choose VOC Pak to produce water that meets the requirement of completely non-volatile organic compounds