Multi functional fluorescent enzyme-linked immunosorbent assay (ELISA) readerApplicable plate types: 6, 12, 24, 48, 96, 384 well plates, standard colorimetric cups, and 12 * 75mm test tubes
Light source: Flash xenon lamp
Detector: 2 photomultiplier tube detectors and 4 photodiode detectors
Oscillation time: 0 to 999 seconds
Temperature control: Room temperature+4 ° C to 60 ° C
Temperature uniformity:< 1°C ,37°C
Temperature accuracy: ±1°C ,37°C
Multi functional fluorescent enzyme-linked immunosorbent assay (ELISA) readerOptical density mode parameters
Wavelength range of 200-1000nm, precise selection of wavelength of 1nm
Bandwidth<=4nm
Wavelength accuracy ± 2.0nm
Wavelength reproducibility ± 0.2nm
Reading range -0.3-4.0 OD
Resolution: 0.001 OD
Measurement accuracy (microporous disk)<± 0.006 OD ± 1.0%, 0-2.0 OD
(Colorimetric dish)<± 0.005 OD ± 1.0%, 0-2.0 OD
Measurement accuracy<± 0.003 OD ± 1.0%, 0-2.0 OD
Scattered light<0.05% @ 230nm
Fluorescence density reading mode parameters
EXWavelength range 250-850nm, increasing by 1nm
EMWavelength range 250-850nm, increasing by 1nm
Bandwidth (Ex, Em) 9nm, 15nm
Accuracy 3fmol/well FITC 200ul in 96 wells
Time-resolved fluorescence mode
Reading direction: top or bottom of microplate
Wavelength range: 250-850 nm
Wavelength selection: increasing by 1nm
Bandwidth (Ex, Em): 9 nm, 15 nm
Accuracy range: Measurement and reading of 1-100 flakes, with a delay setting of 0-600 μ sec before measurement and a time interval of 50-1500 μ sec
Sensitivity: 250 fM europium element (96/384 well plate, top reading)
Fluorescence polarization parameters
Wavelength range: 400-750 nm
Wavelength selectivity: using grating to generate continuous spectrum, increasing by 1nm
Bandwidth (EX, EM): 9 nm, 15 nm
Accuracy:< 5 mP Standard Deviation (1 nM Fluorescein, 96/384 well plate)
Chemiluminescence parameters
Detection method: Color comparison or top and bottom reading of the microplate
Wavelength selection: Full band or individual wavelength selection
Wavelength range: 250-850 nm
Sensitivity:< Firefly luciferase of 2 fg/well (96 well plate, top reading)
Inter hole interference:<0.5% (96/384 orifice plate)