Shanghai Nirun Intelligent Technology Co., Ltd
Home>Products>NTV-CAP1 Cone Plate Rheometer
NTV-CAP1 Cone Plate Rheometer
Technical features of NTV-CAP1 cone plate rheometer: 1.7-inch wide touch screen, easy operation, rich display information. 2. Optional DV data acquisi
Product details

NTV-CAP1 Cone Plate Rheometer

Technical features:

1.7-inch wide touch screen, easy to operate, rich display information

2. Optional DV data acquisition and programmable software can measure rheological properties, and the screen can directly display rheological curves

3. Display shear rate and shear stress indicators

4. Anti static shell and automatic lifting design

5. Built in PT100 temperature probe

6. ARM chip processor, faster data processing speed

7. Durable new bearing design

8. Powerful human-computer interaction interface

9. Gigabit Ethernet port data communication transmission, fast and reliable data transmission.

10. Support external USB data storage; Choose from three storage methods: single point, continuous, and timed.

11. Password protected temperature and viscosity correction coefficients allow users to self correct temperature and viscosity correction coefficients with reliable and accurate data.

12. Supports infinitely variable speed viscosity measurement, and any input speed directly reflects the full range of each rotor, making it convenient for users to select testing conditions.

13. Automatic conversion of dynamic viscosity and kinematic viscosity

14. Arbitrary exchange of multiple viscosity units

15. Multiple user-friendly interface styles are interchangeable

16. Randomly come with powerful operation wizard assistance

17. Continuous variation of viscosity value display

18. There will be an alarm sound when the measurement range is exceeded

19. Full range linear calibration of computers

20. AC power adapter 100V-240V, with good anti-interference performance

Optional attachments:

CAP viscosity standard solution; DV data collection and analysis software

Application:

Low shear rate, measuring low, medium, and high viscosity samples

NTV-CAP1 Main technical indicators: (N: represents rotational speed)

NTV-CAP1 Viscosity range (mPa. s) Measurement accuracy: ± 1% (F.S) Flat temperature range: L: -5 ℃--120℃ H: Room temperature+5 ℃ -250 ℃ Accuracy: ± 0.1℃

model

Rotor number: CPE-40 shear rate: 7.5N sample size: 0.5ml

Rotor number: CPE-41 shear rate: 2.0N sample size: 2.0ml

Rotor number: CPE-42 Shear rate: 3.84N Sample size: 1.0ml

Rotor number: CPE-51 Shear rate: 3.84N Sample size: 0.5ml

Rotor number: CPE-52 Shear rate: 2.0N Sample size: 0.5ml

NTV-CAP1LV

(RPM: 0.1-250)

0.1-3,000

0.5-11,000

0.2-6,000

2-48,000

3-92,000

NTV-CAP1RV

(RPM: 0.1-250)

1-32,000

5-122,000

2-64,000

20-512,000

39-983,000

NTV-CAP1HA

(RPM: 0.1-250)

2.6-65,000

10-245,000

5-128,000

41-1M

78-2M

NTV-CAP1HB

(RPM: 0.1-250)

10.5-261,000

39-982,000

20-512,000

163-4M

314-7.8M

Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!