PAC NSure online total sulfur analyzer
PAC Online Series
Product details

PAC在线总硫分析仪NSure

NSureIntroduction:

NSureOnline sulfur and nitrogen analyzers can be effectively used for continuous monitoring and analysis of sulfur content in refineries, natural gas pipelines, and purification plants, meeting product specifications from the most economical and practical perspective. They have won years of reputation for their stable and reliable performance, simple operation, and no need for too much manual involvement. It is an ideal choice for measuring the sulfur and nitrogen content of natural gas, liquefied petroleum gas, and oil products.

Measurement principle:

Total sulfur analysis:NSure Sulfur

Analyze compliance with standardsASTM D5453, D6667, ISO 20846SH/T0689,GB/T 11060.8

UV Fluorescence Spectroscopy(UVF)Sulfur measurement technology is a fast and accurate method for quantitative analysis of sulfur, which can be used in1-2Complete within minutes as low as20ppbRapid determination of sulfur content up to a hundred levels.

R-S + O2 1000°C SO2 + combustion products

SO2 + hvSO2*SO2 + hν’

A certain amount of sample burns at high temperature, and all sulfur elements are converted into molecular statesSO2After being exposed to ultraviolet light, it emits fluorescence. The generated fluorescence is converted into an electrical signal proportional to the sulfur content in the sample by a photomultiplier tube and detected. Compared to the lead acetate paper tape method for measuring sulfur, the ultraviolet fluorescence method is more stable and does not require consumables, nor does it produce lead waste. This method is not availableXThe issue of matrix interference in radiation can provide accurate analysis results with lower sulfur content without the need for column switching or blowback. Because the detector used eliminates the quenching effects of carbon dioxide and water, you will be able to obtain more stable and excellent linear range analysis results.

Total nitrogen analysis:NSure Nitrogen

Analyze compliance with standardsASTM D4629, DIN 38409, TEIL 27, ASTM D5176, EN 12225

Applying accurate and precise chemical fluorescence detection(CLND)Technology,,Analytical instruments can be used1-2Complete within minutes from20ppbRapid determination of nitrogen content at the hundred level. Chemiluminescence is a rapid nitrogen measurement method that can be used for almost any liquid and gas sample.

R-N + O2 1000°C NO + combustion products

NO + O3NO2* + O2NO2 + O2 + hν’

A certain amount of sample burns at high temperature, and all nitrogen elements are converted intoNOMolecules. thenNOMolecules fully mix with ozone to occur React and emit light. The emitted light is converted into an electrical signal proportional to the nitrogen content present in the sample through a photomultiplier tube and detected.

NSureCharacteristics:

high-precision

Excellent repeatability and precision provide users with high-quality detection data. Good repeatability means that users can fully trust the data it provides. Users can confidently use their data to adjust process conditions to ensure that the total sulfur content in the products produced by desulfurization and blending processes meets the standard. The repeatability of an instrument is measured by its standard deviation.NSureLong term measurements of diesel, gasoline, and natural gas have yielded excellent repeatability results. In most applications, its repeatability over the full range is superior1%The accuracy is better than1%.

rapid analysis

NSureThe analyzer technology is advanced. From sample injection, to complete combustion of the sample, to analysis of the results, the entire process1It can be completed in minutes. Greatly saving time costs. And the analysis and detection process is simple and efficient - no sample residue will be generated

Data interaction interface

Through the touch screen, method settings, calibration, verification, alarm settings, and troubleshooting operations can be directly performed; Equipped with local historical data storage, advanced data output, and interactive functions; Data output control.

Diverse communication methods

4–20 mAIndependent output, digital input and output for remote control and status display

Network Communication ProtocolRTU (Standard configuration)RS-485

Network Communication ProtocolTCP/IPAnd Ethernet(Standard configuration)

Remote Workstation(Optional)

Alarm and Diagnosis

Intelligent design, instruments can perform self diagnosis, facilitating fault handling and repair

NSureTechnical parameters:

Analysis method

NSure Sulfur:UV fluorescence method(UVF)

NSure Nitrogen:chemiluminescent detection(CLND)

Analysis standards

NSure Sulfur: ASTM D5453, D6667, ISO 20846SH/T0689, GB/T 11060.8

NSure Nitrogen: ASTM D4629, DIN 38409, TEIL 27, ASTM D5176, EN 12225

certificate/Regional classification

CSA/UL, NEC Class 1 Division 2 Groups A, B, C & D

ATEX: Zone 1 Ex dbibpxb IIC T4 Gb (T3 with heated valve)

Zone 2 Ex dbnApxb IIC T4 Gb (T3 with heated valve)

Detection range

Sulfur content analysis:

0-1 ppm/wtand0–5% (On site adjustable)

Analysis of nitrogen content:

0-1 ppm/wtand0–2000 ppm/wt (On site adjustable)

Attention: If you need a larger range, please consult the manufacturer.

Flow path

Single channel or dual channel options are available

Detection lower limit(LLD)

20PPb

Company

ppmmg / m3Optional, users need to provide relevant parameters)

Repeatability

less than1%

response time

Each flow path1minute

Analysis time

Each flow path can reach1minute

accuracy

1%

noise

1%

Zero line drift (weekly)

1%

calibration/verification

Use user-defined parameters for automatic or manual system validation and calibration

calibration/Verification time

Single flow path1minute

Communication output

4–20 mAIndependent output

Digital inputs and outputs for remote control and status display

Network Communication ProtocolRTU (Standard configuration)RS-485

Network Communication ProtocolTCP/IPAnd Ethernet(Standard configuration)

Remote Workstation(Optional)

Self diagnostic function

have

Gas requirements

carrier gas:Argon gas, instrument air*Or nitrogen gas(99.975%, 50 psig, 5–15 cc/min

Auxiliary combustion gas: Instrument air*Or oxygenO299.975%), 50 psig ,350-500 cc / min

Pneumatic component air supply: instrument air, 80-100 psig5.6-6.8 Bar

Instrument blowing air supply: Instrument air, pressure of80-120 psig5.6-7 Bar); The starting traffic is425 LPM15 SCFM)The continuous operation flow is130 LPM4.6 SCFM

note appended1Some versions only require instrument air for operation

note appended2Instrument air requires cleanliness, dryness, and no particles

Injection requirements

Sample phase: gas or liquid

Sample inlet pressure: ≤400 psig(depending on the application)

Sample inlet temperature: ≤100? C

Sample conditions: ≤0.5um

Sample return pressure: less than the injection pressure10-20psi keep≥25psiThe pressure returns to the process pipeline

Voltage/electric current/frequency

115 VAC (13 Amps) or 230 VAC (6.5 Amps); 50-60 Hz

power

1500 W

display

LCD(Having7Inch touch screen or15Two types of inch touch screens are available, with touch screens used for programming devices, displaying current and historical analysis results, and alarm status

Temperature requirements

0-40? CUnder extreme conditions, it may be necessary to usevortecCooler (optional)

size

W x D x H :66 x 43.2 x 114.3cm

weight

181.4 Kg

Packaging box size and weight

86” x 42” x 56” 775 lbs.