PW-CTS2-80-40A air-cooled two box type cold and hot shock test chamber (vehicle specification level/vehicle specification Lite level)
1Control methods and characteristics | ||||
Utilizing low-temperature and high-temperature thermal storage tanks,Move the test sample according to the needs of the experiment,And achieve a rapid impact effect,Balanced temperature control system(BTC),useP.I.D.mode controlSSPR,Make the heating amount of the system equal to the heat loss amount,Therefore, it can be used stably for a long time. | ||||
IIperformance:(gascold type,fingerchamberWarm in+25℃,When unloaded) | ||||
1.High temperature slot impact range |
+60℃ ~+150℃ |
|||
2.Low temperature tank impact range |
-40℃ ~-10℃ |
|||
3.Temperature range of high temperature tank |
+60℃ ~+200℃ |
|||
4.Temperature range of low-temperature tank |
-65℃ ~-10℃ |
|||
5.Impact return time |
-40℃ ~+150℃ make an appointment5min no-load(The sensor is placed at the unit air outlet) |
|||
6.temperature uniformity |
±2.0℃ |
|||
7.temperature fluctuation |
±0.5℃ |
|||
8.High or low temperature Impact constant temperature time |
15 minthe above |
|||
9.Basket transfer time |
≦10second(Pneumatic automatic conversion) |
|||
The temperature performance test is based on IEC60068 -3Measurement according to relevant regulations of standards; The sensor is placed at the unit air outlet | ||||
IIIstructure | ||||
1.Studio Size |
W400 ×H400 ×D500 mm |
|||
2.External dimensions |
make an appointmentW1290 ×H 1900 ×D2315 mm |
|||
3.Box Structure |
Having an independent product testing area,High temperature heat storage area,Low temperature storage zone |
|||
4.Inner box material |
Stainless steel plate(SUS 304#) |
|||
5.External material |
Steel plate baking paint |
|||
6.Insulation material |
High temperature tank:24kglass wool Low-temperature tank:PUfoaming+glass wool |
|||
7.heater |
Bare wire typeElectric heater |
|||
8.Supply air circulation system |
a.Dongyuan Motor b.Stainless steel extended axis c.Multi wing fan blade(SIROCCO FAN) |
|||
9.Box door |
onefilm gate,High temperature tank and low temperature tank1type a.Flat embedded handle b.Rear button:SUS 304# c.Silicone foam strip |
|||
10Test hole |
structureψ50mmhole1only,Silicone plug1Only. |
|||
11.Sample placement rack |
stainless steelSUS 304#storage tray2The film. |
|||
12.Mobile wheels |
Including mobile casters (with foot cups). |
|||
13.Transmission mode |
Cylinder transmission. |
|||
14.Tightly pressed |
Silicone foam is tightly packed. |
|||
4、 Freezing system | ||||
1.compressor |
French Taikang fully enclosed compressor . |
|||
2.refrigerant |
Non fluorinated environmentally friendly refrigerantR404A R23,Compliant with environmental regulations,Safe and non-toxic. |
|||
3.condenser |
Automatic heat dissipation high-efficiency finned motor with attached heat dissipation(Air-cooled type). |
|||
4.evaporator |
High performance fin type automatic load capacity adjustment,Can be used for a long time under low temperature and high humidity conditions without frosting. |
|||
5.Other attachments |
Expansion valves, oil separators, desiccants, and other components are all imported from internationally renowned brands. |
|||
6.Refrigerant flow control |
Automatic adjustment of energy consumption output control for refrigeration systems. |
|||
7.Refrigeration process |
Fully implement nitrogen protection welding,Double stage rotary vane pump for vacuum pumping,Ensure the interior of the refrigeration system Clean and reliable. ※ The bottom of the compressor is designed with a water receiving tray,The condensed water is discharged outside the box through the drainage pipe at the back of the box. |
|||
5、 Controller | ||||
1.controller |
Programmable True ColorLCDTouch basedPLCArchitecture Controller Multi group IntelligencePIDcontrol
|
|||
2.Screen display function |
7Inch resolution: 800*480, sixty-five thousand five hundred and thirty-fivetrue color, LEDBacklit display screen Chinese/English language switching display, true color touch input Temperature setting for each box(SV),actual(PV)Value displayed directly; Can display the execution program number、Current segment function、Remaining time and number of cycles,Running time display; Program editing and graphical curve display; Automatic defrosting prompt; Real time display of graphical curves during program execution,Jumpable segments, maintaining functionality; Automatic fault prompt with corresponding solution prompts. |
|||
3.control resolution |
temperature:+ 0.1℃;time: 1min. |
|||
4.Setting Range |
Upper limit of preheating temperature in high temperature chamber:+ two hundred and twenty℃ Lower limit of preheating temperature for low-temperature chamber:- eighty℃ laboratory(Sample area):high temperature+60℃ ~+75℃; low temperature: - ten℃ ~-65℃ |
|||
5.Operation mode |
First, high temperature→ Low temperature Low temperature first→ High temperature |
|||
6.Program capacity |
Available program capacity:maximum127Group; Time setting:each paragraph530hour59Divide; Repeatable command execution:Each command can reach32000Next cycle. |
|||
7.Setting method |
Human computer dialogue mode, using touch input and control. |
|||
8.communication interface |
Can connect to a computer to display curves,Data collection; Can be used as a monitoring and remote control system; Can be used for synchronous control of multiple machines; RS-232、RS-485. |
|||
9.USB storage card |
pluggable1G-32G UDisk DownloadHistorical curve,Historical numbersoccupy,Control system parameters, canHot swappable function. |
|||
10.Data recording method |
Having a beltbatteryprotectiveRAM,Can save the set values, sampling values, and sampling time of the device;The curve recording period can be set30~300 sec,Maximum Memory Time Storage Continuous Storage90Historical curve of the sky,HISTORICAL DATA(When the sampling time is1min),When not used continuously10We have annual data available. |
|||
11.Power off memory function |
The power-off recovery mode can be set to:Hot up/Cold up/stop it. |
|||
12.Appointment startup function |
You can freely set the startup time,After turning on the power, the machine will automatically run. |
|||
13.Software usage environment |
Simplified ChineseWindows2000Or Simplified ChineseWindows XPOperating system. |
|||
sixControl panel | ||||
a.Emergency stop switch b.power switch c.Overtemperature protector d.RS-485interface | ||||
sevenSafety protection device | ||||
a.Heater air burn protection switch b.Heater overcurrent circuit breaker c.Non fuse switch d.Line circuit breaker e.Circulating fan overcurrent overload protection f.Compressor high voltage protection switch g.Compressor overheat protection switch h.Compressor overcurrent protection switch i.Under reverse phase protection switch | ||||
8、 Installation and usage conditions | ||||
1.Use power supply |
AC 3ψ4W 380V 50HZ(R.S.T.NAdd up the grounding wire)(Voltage fluctuation ≤±10%). |
|||
2.compressed air source |
Please ask the customer to provide0.6MPaThe above compressed air sources. |
|||
3.surrounding environment |
Operating within the guaranteed ambient temperature range: five~35℃. |
|||
PS. 1.Please allocate the above power requirements to the terminal block of the machine control box,The customer is required to provide a dedicated non fuse switch for this device; 2.The machine enters the doorway, passage, elevator, etc,Please confirm if you can enter; 3.This quotationExcluding the cost of power cords outside the machine. | ||||
host important material material unmixed one | ||||
order name |
name |
brand |
notes |
|
1 |
compressor |
France Tai Tai |
||
2 |
oil separator |
Emerson, USA |
|
|
3 |
plate heat exchanger |
GermanyGEA |
|
|
4 |
Pressure Switch |
DenmarkDANFOSS |
|
|
5 |
condenser |
Guangzhou Yongqiang |
|
|
6 |
evaporator |
Guangzhou Yongqiang |
|
|
7 |
dryer |
DenmarkDANFOSS |
|
|
8 |
expansion valve |
DenmarkDANFOSS |
|
|
9 |
expansion valve |
AmericaHONEYWELL |
|
|
10 |
solenoid valve |
Castel |
|
|
11 |
controller |
Asusalpha |
China |
|
12 |
Non fuse switch |
Shilin, Taiwan |
|
|
13 |
Circuit breaker protector |
Schneider, France |
|
|
14 |
AC contactor |
Schneider, France |
|
|
15 |
thermal relay |
Schneider, France |
|
|
16 |
Phase sequence relay |
Carlo Gavazzi |
|
|
17 |
Solid state relay |
Carlo Gavazzi |
|
|
18 |
Intermediate relay |
Carlo Gavazzi |
|