Portable methane non methane total hydrocarbon analyzerAdvantages:
Not only can it accelerate the analysis cycle and improve work efficiency, but it also has many advantages such as easy operation, accurate data, and multi-functional one machine.
Characteristics of methane non methane total hydrocarbon analyzer:
1. The instrument has various functions such as power on self-test, power-off protection, automatic restart, automatic ignition, and automatic calibration, ensuring high sensitivity, stability, and repeatability of the entire instrument.
2. The automatic sampling system can be connected to various sample inputs, and the data processing system can be integrated with the data transmission modules of various central control systems to ensure real-time sampling, remote control, real-time transmission, and real-time results.
Portable methane non methane total hydrocarbon analyzerInstructions for use:
(1) Daily inspection of complete system equipment such as analytical instruments, gas processing systems, high-temperature probes, cooling water, and purging gases, including:
1. Check if the instrument displays any fault alarms
2. Observe the changes in sample gas flow rate
3. Operating status of condenser
4. Peristaltic pump drainage or water accumulation in the reservoir
5. Whether the operating status of the analytical instrument and the indicated values of the analytical instrument are normal
6. Heating condition of sampling probe
7. Compressed air pressure for blowing (greater than 0.3MPa) and manual drainage of the front stage air-water separator
8. Does the humidity alarm membrane filter sound an alarm, and does the high-precision filter have water or dust ingress
9. Usage of standard gas
10. Analyze the temperature inside the instrument room
11. Other abnormal situations during operation
(2) Regularly (within one month) perform zero and range calibration on the measured components of analytical instruments, and keep records. When calibrating analytical instruments, the calibration of standard gas cylinders should be accurate.
Application of methane non methane total hydrocarbon analyzer:
Smoke emission detection
Monitoring of trace hydrocarbons in environmental gases
Vehicle and internal combustion engine emission monitoring
Emission boundary monitoring
Detection of trace hydrocarbons in environmental gases
Monitoring of hydrocarbon emissions
Catalytic conversion test
Engine combustion efficiency detection
Monitoring of hydrocarbon pollution in air or other gases