The sensor adopts a closed reference electrode structure, with referenceelectrodeSealed inside a glass electrode to protect the reference electrode from contamination and erosion by certain chemical components in water, such as sulfides, cyanides, chlorides, sulfites, etc. A large capacity dual contact salt bridge can provide sensors with a longer working life and greatly reduce contamination of the reference solution. There are multiple options for Qc32 pH/ORP sensors, which can be selected according to customer needs and operating conditions.
All sensors have many installation methods, including 1/2 ", 3/4", 1 ", 2" NPT, etc. The material is hot-melt PEEK, RYRON and other materials with strong chemical corrosion resistance and high-strength mechanical properties. And designed as a more advanced turbulent self-cleaning structure
Product parameters
Measurement range: 0~14pH- 1000mV~+1000mV
Sensitivity: 0.02pH; 0.2mV
Stability: 0.02pH; 2mV without accumulation
Body material: PEEK,Ryton, Glass Viton, platinum or gold (ORP only)
Temperature compensation: Pt1000RTD
Temperature range: 0-100 degrees
Pressure range: 0-8.0 bar
Maximum flow velocity: 3 meters per second
Installation form: 1/2 "NPT 3/3" NPT 1 "2" plug-in type