Radiation ionization chamber monitoring instrumentICS-323C
This survey instrument has a wide measurement range. It can measure 1cm dose equivalent rate (1 μ Sv/h~300mSv/h) and 1cm cumulative dose equivalent (0.3~10 μ Sv). By opening the shielding cover in front of the ionization chamber, surface contamination of beta rays can be detected. Can be used for radiation detection in facilities for nuclear power, nuclear medicine, radioactive isotopes, accelerators, and X-ray machines.
Radiation ionization chamber monitoring instrumentICS-323CMain features:
1. Small in size and light in weight, easy to carry: weighing only 620g. Comes standard with shoulder straps, wristbands, and handles, suitable for use in various occasions.
2. Automatic zero adjustment: When the power is turned on, the zero adjustment is automatically performed, which is convenient to use.
3. Data storage: It can store up to 3000 measurement data (measurement date, dose rate value, cumulative dose value).
4. Waterproof design: daily waterproof drip structure
5. Compact design and rich functionality: capable of measuring 1cm cumulative dose equivalent (0.3 μ Sv~10 μ Sv); Use commercially available No. 5 batteries; Equipped with backlight illumination, it can be used in dark places; By using the Hold function, the current measured reading can be frozen. When it is difficult to observe the displayed value at certain measurement positions such as deep pipes, this function can be used to hold the measured value of the actual measurement location for easy reading. )
Main technical parameters:
1. Measurement of radiation: X-rays, gamma rays, and beta rays
2. Detector: cylindrical ionization chamber, effective volume of ionization chamber: about 400cm3; β - ray detection window: approximately 4mg/cm2 Maillard film, effective window area: approximately 45cm2; Beta ray shielding cover: thickness approximately 500mg/cm2
3. Energy response: 30keV~2MeV (X-ray, gamma ray): Correction constant ratio relative to 137Cs is 0.85~1.15
4. Measurement range: 1cm Dose equivalent rate: 1 μ Sv/h to 300mSv/h; Accumulated 1cm dose equivalent: 0-10 μ Sv
5. Measurement accuracy: The allowable error of the indicated value is within ± 10% ± 1 digit
6. Display method: Digital display on LCD monitor and analog pointer display
7. Range:
1) Simulation display: 0~10, 100, 1000 μ Sv/h, 10, 100, 1000mSv/h automatic switching (large indication value 300mSv/h); 0〜10μSv
2) Digital display: 0.0~99.9, 100~999 μ Sv/h, 1.00~99.9, 100~300mSv/h; 0.0〜9.9μSv
8. Response time: within 5 seconds
9. Data storage: 3000 data (measurement date, dose rate value, cumulative dose value) can be saved by pressing the MemoRY button
10. Recorder output: DC analog output (0~+100mV/Full) logarithmic 6-digit output (1 μ Sv/h~1000mSv/h)
11. Backlight illumination: Yes (automatic OFF time can be set arbitrarily within 1-10 minutes)
12. Automatic power OFF function: Yes (the automatic OFF time can be set arbitrarily within 0-999 minutes, set to 0 minutes to indicate no automatic power OFF)
13. Battery detection: The LCD display shows the remaining battery level (5 stages)
14. High voltage detection: displayed through the FUNCTION button operation
15. Waterproof performance: JIS protection level IPX2 (equipment tilted within 15 degrees can protect against vertically falling water droplets)
16. Power supply: 4 No. 5 dry batteries
17. Battery life: Over 80 hours of continuous use
18. Temperature and humidity range for use: 0~+40 ℃, below 90% RH (without frosting)
19. Dimensions: Approximately 9 (W) x 17 (D) x 10 (H) cm
20. Weight: Approximately 620g (including dry battery)