product name
|
Rapid changes in temperature chamber
|
application
|
The temperature rapid change test chamber is suitable for testing various performance indicators of aerospace products, information electronic instruments, materials, electrical and electronic products, and various electronic components in high and low temperature or humid and hot environments.
|
Box Structure
|
The temperature rapid change test box body is processed and formed by CNC machine tools, with a beautiful and elegant appearance, and adopts a non reactive handle for easy operation. The inner liner of the box is made of imported high-grade stainless steel (SUS304) mirror panel, and the outer liner of the box is made of A3 steel plate spray painted, which increases the appearance texture and cleanliness. The water replenishment tank is located at the bottom right of the control box and has automatic water shortage protection, making it more convenient for operators to replenish water sources. The large observation window is equipped with lighting to keep the inside of the box bright, and uses tempered glass embedded in the heating body to observe the condition inside the box clearly at any time. Separating the humidification system pipeline from the control circuit board can prevent malfunctions caused by water leakage in the humidification pipeline and improve safety. The waterway system pipeline circuit system adopts a door opening, which is convenient for maintenance and inspection. A double-layer high-temperature resistant high tensile sealing strip is used between the door and the box to ensure the sealing of the testing area. The lead hole (on the left side of the machine) can be used for external testing of power or signal lines (with a diameter of 25mm, 50mm, or 100mm). (The aperture or number of holes must be specified when placing the order). The bottom of the machine adopts high-quality fixed PU movable wheels.
|
refrigeration system
|
Temperature rapid change test chamber refrigeration system and compressor: In order to ensure the cooling rate and minimum temperature requirements of the test chamber, this test chamber adopts a binary cascade air-cooled refrigeration system composed of a Taikang fully enclosed compressor. The cascade refrigeration system consists of a high-temperature refrigeration cycle and a low-temperature refrigeration cycle, with the connecting container being an evaporative condenser. The evaporative condenser also plays a role in energy transfer, transferring the thermal energy inside the working chamber through a two-stage refrigeration system to achieve the goal of high temperature. The design of the refrigeration system applies energy regulation technology, which is an effective way to ensure the normal operation of the refrigeration unit while effectively adjusting the energy consumption and cooling capacity of the refrigeration system, reducing the operating costs and failure rate of the refrigeration system to a more economical state. Refrigerant: Adopting DUPONT R404 (high temperature cycle) and R23 (low temperature cycle) Auxiliary components: expansion valve (Danfoss, Denmark), solenoid valve (CASTEL, Italy); Filter (American Aigo); Refrigeration components such as oil separators (Taiwan Crown Pressure) are all imported.
|
refrigeration principle
|
The high and low refrigeration cycles of the temperature rapid change test chamber both adopt the inverse Carnot cycle, which consists of two isothermal processes and two adiabatic processes. The process is as follows: the refrigerant is adiabatically compressed to a higher pressure by the compressor, consuming work to increase the exhaust temperature. Then, the refrigerant exchanges heat with the surrounding medium through the condenser at an isothermal state, transferring heat to the surrounding medium. After the refrigerant undergoes adiabatic expansion through the valve, the temperature of the refrigerant decreases. Finally, the refrigerant absorbs heat from the object at a higher temperature through the evaporator, causing the temperature of the cooled object to decrease. This cycle repeats itself to achieve the goal of cooling down
|
Compliant with standards
|
This product meets the requirements of GB2423.22-2001 Test N "Temperature Change Test Method"; Meet the requirements of national standards and other relevant standards. Strictly follow GB 10592-89 "Technical Conditions for High and Low Temperature Test Chambers"
|
Product images
|

|
model
|
KWB-100
|
KWB-225
|
KWB-500
|
KWB-800
|
KWB-010
|
Studio size D × W × H
|
450×450×500
|
500×600×750
|
800×700×900
|
800×1000×1000
|
1000×1000×1000
|
performance index
|
temperature range
|
A:-20℃~150℃ B:-40℃~150℃ C:-60℃~130℃ D:-80℃~150℃
|
Humidity range
|
30~98%RH
|
Fluctuations/Uniformity
|
±0.5℃/±2℃
|
Humidity deviation
|
+2、-3%RH
|
heating up time
|
The average temperature rise and fall per minute is 5 ℃ to 20 ℃ (adjustable)
|
Cooling time
|
The average temperature rise and fall per minute is 5 ℃ to 20 ℃ (adjustable)
|
control system
|
controller
|
Imported programmable touch LCD Chinese dialogue display Microcomputer integrated controller
|
Accuracy range
|
Setting accuracy: Temperature 0.1 ℃, Humidity 1% RH, Indication accuracy: Temperature 0.1 ℃, Humidity 1% RH
|
sensor
|
Platinum resistance PT100 Ω/MV
|
Heating system
|
Fully independent system, nickel chromium alloy electric heating heater
|
Humidification system
|
External isolated, all stainless steel shallow surface evaporative humidifier
|
dehumidification system
|
Adopting evaporator coil dew point temperature laminar contact dehumidification method
|
Water supply system
|
The humidification water supply adopts automatic control And it can recycle residual water Water conservation and consumption reduction
|
refrigeration system
|
Fully enclosed air-cooled single-stage compression refrigeration method/Original French "Taikang"/Fully enclosed air-cooled cascade compression refrigeration method
|
circulatory system
|
Temperature resistant and low-noise air conditioning motor Multi blade centrifugal wind turbine
|
Using materials
|
External material
|
High quality carbon steel plate Phosphating electrostatic spraying treatment/SUS304 stainless steel matte surface line texture treatment
|
Inner box material
|
SUS304 stainless steel plate
|
Insulation material
|
Polyurethane rigid foam/ultra-fine glass fiber cotton
|
Door frame insulation
|
Double layer high and low temperature aging resistant silicone rubber door sealing strip
|
Standard configuration
|
1 set of multi-layer heating defrosting with illuminated glass window, 2 sample racks, and 1 test lead hole (25, 50mm)
|
Safety protection
|
Leakage, short circuit, overheating, water shortage, motor overheating, compressor overpressure, overload, overcurrent protection/controller power-off memory
|
supply voltage
|
AC380V ± 10% 50 ± 0.5Hz three-phase five wire system
|
Environmental temperature for use
|
5℃~+30℃ ≤85%RH
|
Note: 1The above data are all at an ambient temperature (QT) of 25 ℃ Measured under no-load conditions in the studio 2. Non standard models can be customized according to specific user requirements“Temperature Rapid Change Laboratory” This technical information is subject to change without prior notice
|
|
|