brand Other brands Origin category domestic
Price range 1-5 thousand Instrument material Steel aluminum wood experimental platform
Instrument structure Single sided experimental platform application area Medical and health care, food, chemical industry, agriculture, geology and mining

SW-CJ-1G single person single-sided (horizontal air supply) purification workbench


SW-CJ-1G single person single-sided purification table (horizontal air supply)This is a purification workbench widely used in medical work, pharmaceuticals, chemical testing, and other fields, providing a sterile and dust-free clean environment.


SW-CJ-1G single person single-sided purification table (horizontal air supply)characteristic

1. Horizontal manifold

2. The quasi closed glass air door can effectively prevent external airflow from entering and the stimulation of human body by operating odors.

3. Adopting an adjustable air volume fan system and remote control switch for electrical appliances. Ensure that the wind speed in the workspace is always in an ideal state.

4. LED LCD control panel operation control

5. The work area is made of 201 stainless steel


SW-CJ-1G structural schematic diagram

technical parameter

model

SW-CJ-1G

Cleanliness

Level 100 @ ≥ 0.5μm(Federal 209E)

cfu

<0.5 pieces/dish Time (90mm diameter petri dish)

noise

≤62dB(A)

Average wind speed

0.25m-0.45/m/s (fast and slow dual speed)

Vibration half peak value

<0.5um (x, y, z direction)

illumination

≥300LX

Power Supply

AC, Single phase 220V/50HZ (200W)

power waste

300W

weight

85KG

Working size

700*400*515mm

External dimensions

855*550*1600mm

Specification and quantity of high-efficiency filters

695*600*38*1

Specification and quantity of fluorescent/ultraviolet lamps

15W*1/15W*1

structure characteristics

The purification workbench consists of several major components such as a box, fan, high-efficiency filter, and operation switch. The box body is made of cold plates and other materials, with a surface spray treatment. The workbench is made of stainless steel. The purification unit adopts a fan system with adjustable air volume. By adjusting the working state of the fan, the average wind speed in the clean working area can be maintained within the rated range, and the service life of the high-efficiency filter can be effectively extended.

working principle

The air passes through the primary filter, is pressed into the static pressure box by the centrifugal fan, and then filtered by the high-efficiency filter before being blown out from the air outlet to form a clean airflow. The clean airflow flows through the area that needs to be purified at a uniform cross-sectional wind speed, carrying away the dust in the area and creating a high cleanliness working environment.

Installation and use

The location of this workbench should be in a relatively clean working room (located in a primary purification room of 100000 or 300000 levels), plugged in the power, and turned on according to the functions shown on the control panel. Before starting up, the working area and shell of the purification workbench should be carefully cleaned to remove surface dust. Normal operation and use can be carried out within ten minutes after starting up.

maintain

1. According to actual usage, regularly remove and clean the primary filter, with a cleaning cycle of generally 3-6 months. If not cleaned for a long time, dust accumulation will affect the insufficient air intake and reduce the cleaning effect

2. When the ideal cross-sectional wind speed cannot be achieved after normal replacement or cleaning of the primary filter air filter, the operating voltage of the fan should be adjusted to achieve the ideal uniform wind speed.

3. Generally, when the working voltage of the fan is adjusted to the desired level after 18 days of use, and the ideal wind speed cannot be achieved, it indicates that the high-efficiency filter has accumulated too much dust (the filter holes on the filter material are basically blocked and need to be updated in a timely manner). The service life of a high-efficiency air filter is generally 18 months.

4. When replacing the high-efficiency air filter, attention should be paid to the correctness of the model, specifications, and dimensions (original configuration), install the device according to the direction of the arrow, and pay attention to the sealing around the filter to ensure no leakage occurs.

General faultsReasons and elimination methods

Fault phenomenon

reason

Discharge method

The main power switch cannot be closed and automatically trips

1. Fan jamming causes motor stalling or short circuit in the circuit

1. Adjust the position of the fan shaft, or replace the impeller and bearings, and check if the circuit is intact.

2. Check the insulation resistance of the circuit and components to the casing point by point according to the circuit diagram, and repair the insulation fault.

Low wind speed

1. Excessive dust accumulation in the primary filter.

2. The high-efficiency filter has failed.

1. Clean the primary filter.

2. Replace the high-efficiency filter.

The fan is not turning

1. The contactor is not working.

2. The fuse of the fan power supply has blown.

1. Check if the contactor circuit is normal.

2. Replace the melt core.

The fluorescent light is not on

1. The lamp or relay is damaged.

2. The fuse of the lamp power supply has blown.

1. Replace the light tube or relay.

2. Replace the melt core.