SanfengMitutoyoGrating ruler (optical ruler) maintenance, debugging, etc
MitutoyoCommon faults and maintenance of Sanfeng grating ruler
Not displaying after connecting the power supply
(1)Check for broken wires and good plug contact;
(2)The power fuse is blown properly;
(3)Whether the power supply voltage meets the requirements;
(4)Check if the connection section with the grating ruler is in good contact;
(5)Recheck if the installation of the reading head meets the installation requirements;
(6)Check if the output signal of the grating ruler is normal
Inaccurate readings/Do not read numbers/System alarm
(1)Check if the grating ruler is installed correctly and if the fixing screws are loose
(2)Is the plug in good contact;
(3)Does the grating ruler collide with other components when moving
(4)Check if the precision of the machine tool guide rail motion pair is too low, causing changes in the working gap of the grating
It consists of two main parts: a grating ruler and a position control board. The grating ruler detects the actual displacement of the machine tool and outputs two signals related to the displacement amount and direction to the position control board of the CNC system, forming a fully closed-loop control system
If the above treatment still cannot eliminate the fault, please contact the agent or manufacturer. Do not disassemble by yourself.
AT-112Sanfeng optical ruler measuring optical projector/pixiv/CNCSpecialized grating ruler for type instruments
AT103Grating rulerAT115Grating ruler 539series
AT211A / AT211BGrating ruler
539Series AT715 grating ruler
529seriesAT11-TLGrating ruler|Optical ruler
AT102Grating ruler/Optical ruler
174-173B/174-175Bmonitor
DROThe system counter is compatible with milling and lathe counters
AT211Slender grating rulerDirectly connected to the controller, using a parameter selection method that customizes the grating ruler according to the expected usage purpose.
Can provide wide range resolution: 0.1 μ m~5mm and response speed up to 120m/min