TP-7700 gas chromatograph
The TP-7700 gas chromatograph is a simple gas chromatograph with complete functions, stable performance, reasonable structure, durability and reliabil
Product details
main features
1、 Stable and reliable column box system
1. The first-order program temperature can be manually controlled to meet different analysis needs.
2. Overheating protection function: When the temperature of the column box exceeds 420 ℃, the power will be automatically cut off
2、 Flexible sampler system
1. Fill column sampler, install packed column to meet daily chromatographic analysis work
2. Large caliber capillary column sampler for complex sample analysis
3. Gas injection valve: specifically designed for the analysis of gas samples
4. The glass column system is particularly suitable for pesticide analysis
5. Sample injector overheat protection: automatically cuts off power supply if temperature is abnormal
3、 Simultaneously install FID and TCD detectors
4、 File storage function
1. Users can save/read commonly used settings such as aging, secure shutdown, and similar analysis as needed
2. Storage of 6 analysis condition parameters
Technical parameters
● Column box
Temperature control range: 30 ℃ to 399 ℃ at room temperature
Temperature control accuracy: ± 0.1 ℃
Overheating protection function: When the temperature of the column box exceeds 420 ℃, the power will be automatically cut off
● Program heating up
Order: 1
Heating rate: 0.1-39.9 ℃/min, when<200 ℃
0.1-25 ℃/min, ≥ 200 ℃
Total program time: 65535 minutes
● Detector
Temperature control range: 30 ℃ to 399 ℃ at room temperature
Temperature control accuracy: ± 0.1 ℃
Overheating protection function: automatically cuts off power when temperature is abnormal
(1) Flame ionization detector (FID)
Structure: Cylindrical collection electrode, metal nozzle, easy to install
Detection limit: ≤ 5 × 10-11g/s (n-hexadecane)
Baseline noise: ≤ 5×10-14A
Baseline drift: ≤ 2×10-13A/30min
(2) Thermal conductivity detector (TCD)
Structure: Semi diffusive, four arm rhenium tungsten wire
Sensitivity: ≥ 2000mV · ml/mg (n-hexadecane)
Baseline noise: ≤ 0.03mV
Baseline drift: ≤ 0.1mV/30min
Start time: ≤ 2 hours