1TW-6636 online ammonia analyzerbrief introduction
TW-6636The online ammonia analyzer is a high-precision online automated instrument with functions such as automatic completion of chemical reactions, optical detection, graphic and textual display, control output, and data storage; The instrument adopts a liquid crystal display, which displays measurement results, system information, and a fully Chinese menu operation interface in rich text, charts, and curves; The combination of humanized design concept and high-tech has achieved the performance and product competitiveness of this instrument.
2TW-6636 online ammonia analyzermain features
1) Advanced embedded microcontroller technology;
2) Exquisite structure, disc installation, all aluminum frame box, beautiful and sturdy, strong anti-interference ability;
3) Large screen dot matrix LCD with intuitive and rich display content, capable of switching between single and multiple channels;
4) Adopting the principle of constant pressure dosing, the structure is simple, the measurement accuracy is high, and maintenance free;
5) Adopting 450nm imported light source and photovoltaic cell, with long lifespan, small drift, stability, and reliability;
6) Programmable multi-channel selection design, capable of modifying channel measurement cycles, effectively saving reagents;
7) Equipped with temperature measurement function, it can compensate for measurement data based on temperature;
8) The specific black box function can query historical data, operation records, and calibration records;
3Main technical performance
1) Range: (0-100) µ g/L
2) Display: 320 × 240 dot matrix LCD, Chinese menu
3) Measurement cycle: 5 minutes shorter than usual
4) Instrument indication error: ± 2% F.S
5) Repeatability: not greater than 1%
6) Stability: Baseline drift: using blank calibration, no impact on measurement
Chemical drift: ± 1% F.S/24h (varies depending on reagent stability)
7) Sample conditions: Flow rate: (150~300)mL/min
Temperature: (5-50) ℃
Pressure: (14~140)KPa
The allowed solid content of the water sample is not greater than 5 microns (no gel like substances are allowed)
8) Environmental temperature: (5-45) ℃
9) Environmental humidity: not exceeding 90% RH (without condensation)
10) Reagent type: 1 type
11) Reagent consumption: The longer the measurement cycle, the less reagent consumption;
12) Isolated output: (4-20) mA (isolated output, one per channel)
13) Power supply: AC (85-265) V, frequency (45-65) Hz
14) Power: 60W
15) Dimensions: 690mm x 450mm x 300mm (height x length x depth)
16) Hole size: 645mm × 410mm
17) Weight: 25kg
18) Alarm: Sample breakage alarm, over limit alarm