BCM-1600Type biological purification workbenchinstructions
Three person biological purification workbenchProduct Features
1Full circulation airflow design, when there is a large amount of fresh air supply in the work area, the exhaust outlet is turned on.
2The workbench is equipped with up and down movable doors in front and behind, providing flexible and convenient positioning.
3Air curtain isolation design to prevent internal and external cross contamination.
4The adjustable air volume fan system is adopted, and the electrical appliances are controlled by LCD switches to ensure that the air speed in the working area is always in the ideal state.
Three person biological purification workbenchtechnical parameter
model parameter |
BCM-1600 |
|
Cleanliness level |
100level@≥0.5μM(Federal Republic of the United States)209E) |
|
cfu |
≤0.5individual/Dish · hour (Φ)90mmCultivate petri dishes |
|
Average wind speed |
0.25~0.45m/s |
|
noise |
≤62dB(A) |
|
Vibration half peak value |
≤0.5μM(x、y、zDirection) |
|
illumination |
≥450Lx |
|
Power Supply |
ACsingle-phase220V/50Hz |
|
maximum power |
800W |
|
weight |
200Kg |
|
Size of workspace |
W1×D1×H1 |
1600×700×620 |
External dimensions |
W×D×H |
1800×740×1950 |
Applicable number of people |
threepeople/Single sided |
|
Specification and quantity of high-efficiency filters |
1795×640×50×① |
|
Fluorescent lamp/Specification and quantity of ultraviolet lamps |
30W×①/30W×① |
structure characteristics
The purification workbench consists of several major components such as a box, fan, high-efficiency filter, and operation switch. The box body is made of cold plates and other materials, with a surface spray treatment. The workbench is made of stainless steel. The purification unit adopts a fan system with adjustable air volume. By adjusting the working state of the fan, the average wind speed in the clean working area can be maintained within the rated range, and the service life of the high-efficiency filter can be effectively extended.
working principle
The air passes through the primary filter, is pressed into the static pressure box by the centrifugal fan, and then filtered by the high-efficiency filter before being blown out from the air outlet to form a clean airflow. The clean airflow flows through the area that needs to be purified at a uniform cross-sectional wind speed, carrying away the dust in the area and creating a high cleanliness working environment.
Installation and use
The workstation should be located in a relatively clean working environment (preferably in a primary purification room of 100000 or 300000 levels), plugged in the power, and turned on according to the functions shown on the control panel. Before starting up, the working area and shell of the purification workstation should be thoroughly cleaned to remove surface dust. Normal operation and use can be carried out within ten minutes after starting up.
maintain
1According to actual usage, regularly remove and clean the primary filter, and the cleaning cycle is generally3-6A month. If not cleaned for a long time, dust accumulation will affect the insufficient air intake and reduce the cleaning effect
2When the ideal cross-sectional wind speed cannot be achieved after normal replacement or cleaning of the primary filter air filter, the operating voltage of the fan should be adjusted to achieve the ideal uniform wind speed.
3Generally, when the operating voltage of the fan is adjusted to the highest point after 18 days of use and the ideal wind speed cannot be achieved, it indicates that the high-efficiency filter has accumulated too much dust (the filter holes on the filter material are basically blocked and need to be updated in a timely manner). The service life of a high-efficiency air filter is generally 18 months.
4When replacing the high-efficiency air filter, attention should be paid to the correctness of the model, specifications, and dimensions (as configured by the original manufacturer), install the device according to the direction of the arrow, and pay attention to the sealing around the filter to ensure that there is absolutely no leakage.
General faultsReasons and elimination methods
Fault phenomenon |
reason |
Discharge method |
The main power switch cannot be closed and automatically trips |
1The fan is stuck, causing the motor to stall, or there is a short circuit in the circuit |
1Adjust the position of the fan shaft, or replace the impeller and bearings, and check if the circuit is intact. 2Check the insulation resistance of the circuit and components to the casing point by point according to the circuit diagram, and repair insulation faults. |
Low wind speed |
1The initial filter has accumulated too much dust. 2The high-efficiency filter has failed. |
1Clean the primary filter. 2Replace the high-efficiency filter. |
The fan is not turning |
1The contactor is not working. 2The fuse of the fan power supply has blown. |
1Check if the contactor circuit is normal. 2Replace the melt core. |
The fluorescent light is not on |
1The light tube or relay is damaged. 2The fuse of the lamp power supply has blown. |
1Replace the light tube or relay. 2Replace the melt core. |