The YC9100-1 touch based COD rapid analyzer is designed and developed based on relevant national standards. It features a 7-inch full touch screen, flat icon design, a guided operating system, and a simple and easy to understand interface, allowing operators to use it quickly and accurately.
The detection system adopts an independent multi module optical (imported) design with excellent optical performance. The operation does not require manual intervention, and the system automatically identifies the detection channel for precise detection. The detection adopts a 16mm digestion detection integrated optical colorimetric tube, which simplifies the operation steps, avoids errors caused by repeated liquid removal, and greatly improves accuracy.
Technical Parameter
Instrument model | YC9100-1Touch basedCOD rapid analyzer |
Testing items |
COD (Chemical Oxygen Demand) |
Testing standards |
HJ/T399-2007 |
Range of measurement |
0-15000mg/L |
limit of detection |
5mg/L |
Dissolve the environment |
165℃,20min |
Measurement Error |
≤± 5% |
repeatability |
≤±3% |
Optical stability |
≤±0.001A/20Minutes(100000Hour without decay) |
Colorimetric method |
Colorimetric tube (digestion colorimetric integrated tube) |
data processing |
3Ten thousand records96Curve (supports custom extension projects) |
transmission mode |
USBLocal Area Network |
printer |
Built in thermal printer |
Operation interface |
Touchscreen Chinese/English (optional) |
display |
7Inch color touch screen |
Power supply mode |
AC(220V±10%),50Hz |
power |
5W |
work environment |
5-45℃ ≤85%No condensation |
Host size |
348mm*250mm*145mm |
Instrument weight |
host<2.6kg; Dissolver<7kg |
Technical parameters of digestion instrument
Dissolve sample size |
25branch |
time display |
Real time date and time display |
Operation interface |
Chinese display |
operating mode |
Touch-tone |
rated voltage |
AC 220V±10%/50Hz |
Power requirements |
grounding |
Dissolve pore size/depth |
16mm/65mm |
Dissolving capacity |
>12mL |
display screen |
TFTHigh definition color screen |
timing error |
10min±0.1s |
Pre set digestion mode |
CODTotal phosphorus, total nitrogen |
Dissolution method |
Fully enclosed type |
safety protection |
Double protection against overheating |
temperature control |
intelligencePIDtemperature control |
custom mode |
5The temperature and time can be set arbitrarily |
Project switching |
Quick switch |
Reminder method |
buzzer |
Alarm prompt1 |
Reaching the specified temperature and maintaining a constant temperature |
Custom time |
5-999minadjustable |
Alarm prompt2 |
Dissolve completed and stop heating |
Custom Temperature |
90~200℃ adjustable |
Dissolve protection |
Heat shield/shield |
Temp Accuracy |
<±0.5℃ |
Temperature control sensor |
Imported PlatinumPT100 |
heating rate |
Set to165℃,12Constant within minutes |
Instrument weight |
6kg |
work environment |
-10-40℃, relative humidity<85% |
External dimensions |
230*340*130(mm) |
Example of operating interface
Instrument characteristics
1. Full touch design: 7-inch touch screen, supporting Chinese/English display (optional), guided operating system, simple and time-saving.
2. Wide detection range: dozens of detection parameters can be expanded, and customized according to customer needs.
3. Detection optics: Adopting independent multi module optical (imported) design, excellent optical performance, stable and reliable.
4. Operating System: Guided operating system with a simple and easy to understand interface, guiding users to complete operations quickly and accurately.
5. Operating program: a. Built in system program and user program, which can add frequently tested items to the user program for easy searching.
b. It has power-off protection function and one click factory reset function to prevent data loss.
c. The built-in standard work program can be used directly, and users can also expand the detection items themselves.
d. Built in help interface, which can query user detection process problems and issues.
e. It has a one click factory reset function, which can quickly recover data loss in case of unexpected situations.
6. Intelligent detection: Standard measurement and continuous measurement can be switched freely, reducing repeated measurement steps and improving detection efficiency.
7. Operation safety: Adopting a digestion colorimetric integrated tube, reducing the steps of transferring reagents, making the measurement simple, fast, and safe.
8. Quick printing: Equipped with a built-in thermal printer, it can print both current and historical measurement data.
9. Data storage: Standard working curve and 96 user-defined items, capable of storing 30000 pieces of detection data.
10. Online function: Supports data transfer between network port, LAN, USB and PC or printer.
11. Professional reagents: specialized reagents, simplified operating procedures, reduced operational errors, less reagent usage, and lower consumable prices.