1、 Product Introduction:
Ultrasonic water bath oscillatorIt is a biochemical instrument that combines a temperature controlled constant temperature box and a speed regulating multi-purpose oscillator, and has ultrasonic cleaning function. It is mainly suitable for oscillation cultivation of various liquid and solid compounds such as biology, biochemistry, cells, and bacterial strains in various universities, medical, petrochemical, health and epidemic prevention, environmental monitoring and other scientific research departments.During the experimental process, it is often necessary to constantly shake the test solution to prevent precipitation or to compare two or more test solutions at a certain shaking speed and temperature. Using this instrument can achieve ideal results.
2、 Characteristics:
1Adopting a microcomputer temperature controller, the temperature control is precise and the operation is simple.
2Equipped with over temperature alarm and over temperature stop heating function to ensure experimental safety.
3The inner liner is made of high-quality stainless steel, which does not rust.
4Adopting advanced speed control system, the speed runs smoothly.
5High power DC motor ensures long-term continuous operation.
6Temperature timing and speed timing can be controlled separately.
7The ultrasonic power can be adjusted as needed.
IIITechnical parameters:
1Temperature control range: room temperature -100.0 ℃.
2Temperature control method: microcomputer PID control, PT100 sensor.
3Temperature control accuracy: 0.1 ℃.
4Temperature fluctuation: 0.2 ℃.
5Working size: 490 × 390 × 170mm
6Speed range: Start at -300 revolutions per minute.
7Vibration mode: round-trip
8Vibration amplitude: 20mm
8Timing range; 0-999 minutes
9Heating power: 1500W
10Power supply: 220V 50Hz
11Ultrasonic power: 0-600W adjustable
12Number of ultrasonic transducers: 12
13Ultrasonic frequency: 28K
4、 Usage:
1First, add water to the sink, then bottle the reagent onto the shaker and clamp it with a spring.
2Connect the power supply, turn on the power switch, and the power switch light will turn on.
3Controller operation method:
- Click the temperature button, then click △ or ▽ to select the desired working temperature. Once set, click the temperature button again and the set temperature will be saved.
- Click the speed button, then click △ or ▽ to select the oscillation speed that needs to work. After setting, click the speed button again, and the set speed will be saved.
- Click the time button, then click △ or ▽ to select the desired working time. Once set, click the time button again and the set time will be saved. If timing is not required, set the time to 0.
- Press and hold the start stop button for 3 seconds to start running. To stop, press and hold the start stop button for 3 seconds.
4Turn on the oscillation switch, the ultrasound starts running, click the ultrasound up and down button to select the ultrasound power, and use the time button to determine the duration of the ultrasound operation for the player.
5Ultrasonic water bath oscillatormatters needing attention:
1All technical indicators of this series of products have undergone actual operation inspection before leaving the factory. If the user needs to use a self provided thermometer (with certain accuracy) for measurement due to process requirements, the user's thermometer shall prevail. If there is a difference between the displayed value of the temperature control instrument and the thermometer, the user can adjust the internal parameters of the temperature control instrument slightly to meet the process requirements.
2. When using, water must be added first and then electricity must be turned on. Dry burning is strictly prohibited.
3. When in use, there must be a reliable grounding to ensure safety.
4. If the water level is lower than the electric heating tube, it is not allowed to use it with electricity to prevent the electric heating tube from bursting and being damaged.
5. The water level should not be too high to prevent water from overflowing into the electrical box and damaging components.
6. When not in use for a long time, the water in the sink should be drained and wiped clean, and the scale in the sink should be regularly removed.