UniBest SafeLight mini handheld luminescent bacterial comprehensive toxicity analyzer from the United States
UniBest SafeLight mini handheld luminescent bacterial comprehensive toxicity analyzer from the United States
Product details


Brief introduction

In the field of water quality testing, especially in response to emergencies, conventional analytical methods require complex and expensive instruments, long testing times, and single analysis items, which cannot provide fast, accurate, and reliable analysis reports for emergencies. AmericaUniBestThe group company provides you with a toxicity analyzer for rapid detection of water quality conditions. You can quickly assess the toxicity of water quality and provide scientific and reliable analysis reports with just a few simple steps.

SafeLight miniHandheld ultra portable biological toxicity analyzer, combined with the United StatesUniBestThe group company has many years of experience in producing chemiluminescence instruments. It has been widely used in environmental emergency monitoring, disease control centers, water quality and safety monitoring of water companies, public security anti-terrorism emergency monitoring, food safety emergency monitoring, and water quality and safety assurance of field armies, and has obtained highly sensitive and stable reliable detection results.

2、 Working principle

SafeLight miniThe handheld ultra portable biotoxicity analyzer is a toxicity detection system based on biosensing technology, which qualitatively and quantitatively detects the changes in luminescence intensity of luminescent bacteria during metabolism. This instrument uses luminescent bacterial freeze-drying reagents. When these bacteria are in a toxic environment, their emitted light is inhibited, and the toxicity of the sample can be quickly and accurately tested based on changes in their luminescence intensity. It can directly detect thousands of potential toxic substances.

IIImain features

1 Upper limit detection time:15second-15Minutes, duration can be freely set

2 Types of detectable toxic substances:>2500species

3 Used for total toxicity testing(ATPContent)

4 Adopting complianceISO11348-3:20082008-12-31)Marine luminescent bacteria specified by standards; The marine luminescent bacteria also meet the national standardsGB15441-1995For positive quality controlHgCl2Testing requirements; It can also be paired with Vibrio cholerae from Qinghai

5 The device can detect the total amount of microorganisms, with a lower detection limit of:100 cfu/mL

6 High detection accuracy (lower detection limit)10-15molATP

7 Ultra portable, small size, miniature handheld (weight)<300g

8 Ultra low power consumption(2jointAABattery, typical working time exceeds200Hour)

9 Easy to operate (few buttons, user-friendly interface)

10Control intelligent networking, data can be transmitted to computer networks

11 Detection sensitivity: superior to fish body96Acute biotoxicity test method for hourly cultivation determination

4、 Measurement process

5、 Product Features

1 Using internationally advanced biotechnology to detect specific luminescent reactions of toxic substances using luminescent bacteria

2 One machine dual-use, portable or laboratory use

3 High sensitivity, able to detect belowppbLevel of toxic substance content

4 Good accuracy, easy operation, and low cost

5 Fast speed,15second~15The test results can be obtained in minutes

6 Multiple forms of sample tubes can be used

7 The reaction temperature is15~30℃Convenient for use in various occasions

8 Simple data processing, easy to useExcelimmediately

9 The detection range is wide and can detect heavy metal ions including chromium, copper, lead, nickel, mercury, etc,DDT(DDT, a highly effective organic insecticide)PCDs(Dichlorobiphenyl), organic phosphorus2,4-D(Growth agent), cyanide, dioxin, botulinum toxin, chloroform, detergent, diesel Numerous organic and inorganic toxic substances such as fluoranthene. The instrument has good spectral performance and a wide range of applications.

10Capable of self calibration and background analysis (power on)60Seconds) to adapt to changes in ambient temperature

11, can be setNon toxic or low toxicity/poisoning/Heavy Poison/high toxic/Highly toxiccritical value

12Display: LCD display

13Printer: viaRS232External interface

14Input: Use cursor keys to input data

15Consumables: luminescent bacteria and supporting reagent set

6、 Technical indicators

1 Detector: High sensitivity photoelectric cell

2 Detection method: photoelectric counting method

3 Range:0-107cfu/mL

4 Detection sensitivity: (ATP<10-15mol

5 Detection range:0-9999RLUs

6 Dynamic range:4An order of magnitude(0-9999RLUs

7 Detection accuracy:1RLU

8 Data Memory: Relative Luminescence(RLUValue)

9 Correlation coefficient:≥99.5%

10Relevant significance:P<0.01(Pure compound solution)perhapsP<0.05(industrial waste water)

11Detection error:≤5%

12Storage capacity:2000A test result value

13Limit the number of groups:≥100group

14Detecting repeatability:CV≤8%

15Detection interval:15second

16Serial interface:RS-232compatible

17Battery and Capacity:2jointAABattery,2600mAhalkaline

18Battery standby time:6Month

19Battery working time:1000Next test

20Operating temperature range:5-40℃/Relative humidity range:20-85%

21Storage temperature range:-10-40℃/Relative humidity range:20-95%

22Data display/Output:RLUCFUECIC

23Instrument size:75×180×50mmW×H×D

24Sample tube:12×155mm

25Instrument weight: approximately300g

7、 Application Fields

Suitable for water companies, environmental monitoring stations, disease control centers, food and pharmaceutical research institutions, and other units. Whether you are at the testing site or in the laboratory, this instrument is convenient to use and can save measurement data for your future work.

8Technical Parameter

SafeLight miniSensitivity of detection for common toxic pollutants

Toxic Substances

IC50(mg/L)

LDL(mg/L)

Hg

0.04

0.005

Cd

0.07

0.024

Pb

0.1

0.04

Se

1.2

0.3

Cu

0.11

0.01

Cr

0.1

0.03

2,4,5-T

0.05

0.01

2,4-dinitrophenol

0.1

0.05

DDT

0.07

0.01

cyanide

0.45

0.1

2,4,- D

0.05

0.01

parathion

0.1

0.03

IC50 Semi inhibitory concentration, which refers to the production of toxic substances50%The concentration at the time of light loss,Inhibition Concentration

LDL –Lower detection limit,Lowest Detection Limit