Product Usage:
VOCThe release volume environmental testing chamber is mainly used for studying artificial boards and related light industrial productsVOCChange pattern of release amount, detecting in artificial boards and related light industrial productsVOCRelease volume and classification of testing product levels; Comprehensive environmental indicator evaluation of indoor air quality based on indoor pollutant data or predictive models.
equipmentcharacteristic:
1There are mainly insulation boxes, mirror stainless steel testing chambers, clean constant temperature and humidity air supply systems, air circulation devices, air exchange devices, temperature control units for testing chambers, signal control and processing parts (temperature, humidity, flow rate, replacement rate, etc.).The inner chamber is a mirror stainless steel test chamber, and the outer layer is an insulated box, which is compact, clean, efficient, and energy-saving. It not only reduces energy consumption but also reduces the equipment's balancing time.
2The touch control screen is used as the dialogue interface for personnel to operate the equipment, which can directly set and digitally display the temperature, relative humidity, temperature compensation, dew point compensation, dew point deviation, and temperature deviation inside the box.
3. The equipment adopts the internationally advanced jacket structure, which is compact, clean, efficient, and energy-saving
4The device integrates high-definition clean air treatment and humidity regulation, with a compact, efficient, and energy-saving structure.
5The equipment is equipped with full protection devices and system safety operation protection devices, making the operation of the equipment more reliable and safe.6. Imported humidity, temperature and humidity sensors with high accuracy and stable performance.
7. Imported refrigeration unit with stable operation and long service life.
8. The compressor also has overheat, overcurrent, and overpressure protection measures, ensuring safe and reliable operation of the entire machine.
9. The laboratory can be customized according to customer requirements and the size of the studio.
Technical parameters:
temperature
Adjustment range
15℃~40℃
adjusting accuracy
±0.5℃
relative humidity
Adjustment range
(40~80)%RH.
adjusting accuracy
±3%R.H.
Air exchange rate
Adjustment range
0.2m3~2m3/h(air change rate0.2~2order/Hour)
adjusting accuracy
Ventilation rate ±1%
Face Velocity
The air level velocity is maintained at(0.1~2)m/sAccuracy ≤0.05m/s
pressure
10±5Pa
background concentration
Formaldehyde background concentration<0.006mg/m3,TVOCBackground concentration<0.02mg/m3. Single itemVOCBackground concentration<0.002mg/m3
Meet the standards
GB18580—2017、GB18584—2001、GB18587—2001、GB17657—2013